Body composition measures assessed by bioelectrical impedance analysis and dual-energy X-ray absorptiometry in a sample of Brazilian adults and older adults
Frontiers in Nutrition · 06.01.2026 · 12
DOI: 10.3389/fnut.2025.1689031 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Nutrition, online erschienen am 06.01.2026. Verfasst von Vivian Wahrlich und 5 weiteren Autor:innen. Beteiligte Einrichtung: Laboratório de Avaliação Nutricional e Funcional, Departamento de Nutrição Social, Universidade Federal Fluminense, Niterói, Rio de Janeiro, Brazil. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Themen: Gerontologische Pflege
Abstract (Originalsprache)
Introduction
Bioelectrical impedance analysis (BIA) is a common technique for assessing body composition in clinical and epidemiological settings. However, its accuracy is limited compared to reference methods such as dual-energy X-ray absorptiometry (DXA).
Purpose
This study aimed to evaluate the agreement between fat-free mass (FFM) and fat mass (FM) measured using BIA (Tanita BC-418) and DXA and to develop a calibration model to correct BIA estimates in a heterogeneous sample of Brazilian adults and older adults.
Methods
We analyzed data from 945 participants (aged≥18 years; 611 female participants) who underwent both BIA and DXA assessments across multiple cross-sectional research projects. Agreement between the BIA and DXA measures of FFM ( BIA FFM and DXA FFM) and fat mass (FM) was assessed using Pearson correlation coefficients (r) to evaluate precision and Lin’s concordance correlation coefficients (CCCs) to evaluate accuracy. Mean absolute and relative differences were evaluated using paired t -tests or analysis of variance (ANOVA) by sex, age, and nutritional status based on body mass index (BMI). Linear regression was employed to calibrate BIA FFM against DXA FFM. A multivariate prediction model for DXA FFM was developed using BIA-derived resistance, stature, body mass (BM), and age in a randomly selected subsample comprising 70% of the participants ( n = 659) and was validated in the remaining 30% ( n = 286).
Results
BIA and DXA measures were highly correlated for both FFM and FM ( r = 0.97) and demonstrated moderate to high accuracy (CCC ≥ 0.93). For the entire sample, BIA overestimated FFM by 3.1 kg (SD = 2.4; +7.2%) and underestimated FM by 2.9 kg (2.3; −13.0%) compared to DXA (both p < 0.0001). The resulting calibration equation for FFM was DXA FFM = 0.94420 × BIA FFM–(0.01128 x Age) + 0.20516. The multivariate prediction equation derived from the development group was as follows: FFM (kg) = (Sex × 4.1797) + [Stature (cm) × 0.1062] + [Resistance Index (cm 2 / Ω ) × 0.5289] + [Body Mass (kg) × 0.1797] – [Age (yrs) × 0.0705] – 5.4286 (female participants = 0, male participants = 1). In the validation group, the mean FFM values obtained by the calibrated regression and by the new multivariate equation showed no statistically significant difference from the actual DXA FFM measurement.
Conclusion
Significant discrepancies existed between BIA- and DXA-derived body composition measures in this heterogeneous sample of Brazilian individuals. The developed prediction equations effectively calibrated BIA FFM estimates to align with DXA values, providing a practical method to enhance the accuracy of BIA for body composition assessment in this population.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Laboratório de Avaliação Nutricional e Funcional, Departamento de Nutrição Social, Universidade Federal Fluminense, Niterói, Rio de Janeiro, Brazil.
- Programa de Pós-Graduação em Ciências da Nutrição, Universidade Federal Fluminense, Rio de Janeiro, Brazil.
- Programa de Pós-Graduação em Saúde Coletiva, Universidade Federal Fluminense, Rio de Janeiro, Brazil.
- Departamento de Epidemiologia e Bioestatística, Instituto de Saúde Coletiva, Programa de Pós-Graduação em Saúde Coletiva, Universidade Federal Fluminense, Rio de Janeiro, Brazil.
Zitieren
APA
Wahrlich, V., Ciafrino, A., Chain, A., Bossan, F. M., Baltar, V. T., & Dos Anjos, L. A. (2026). Body composition measures assessed by bioelectrical impedance analysis and dual-energy X-ray absorptiometry in a sample of Brazilian adults and older adults. Frontiers in Nutrition, 12. https://doi.org/10.3389/fnut.2025.1689031
Vancouver
Wahrlich V, Ciafrino A, Chain A, Bossan FM, Baltar VT, Dos Anjos LA. Body composition measures assessed by bioelectrical impedance analysis and dual-energy X-ray absorptiometry in a sample of Brazilian adults and older adults. Frontiers in Nutrition. 2026;12. doi:10.3389/fnut.2025.1689031
BibTeX
@article{Wahrlich2026,
author = {Wahrlich, Vivian and Ciafrino, Agnes and Chain, Amina and Bossan, Francine Moreira and Baltar, Valéria Troncoso and Dos Anjos, Luiz Antonio},
title = {Body composition measures assessed by bioelectrical impedance analysis and dual-energy X-ray absorptiometry in a sample of Brazilian adults and older adults},
journal = {Frontiers in Nutrition},
year = {2026},
volume = {12},
doi = {10.3389/fnut.2025.1689031}
}
EUVIMED