Comparison of anterior segment parameter measurements between Sirius Plus and Pentacam AXL
Frontiers in Medicine · 21.04.2026 · 13
DOI: 10.3389/fmed.2026.1819141 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Medicine, online erschienen am 21.04.2026. Verfasst von Jiliang Ning und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, The Third People's Hospital of Dalian, Dalian, China. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
This cross-sectional study compared the differences, consistencies, and correlations of the measurements of anterior segment biological parameters between Sirius Plus and Pentacam AXL in a healthy population with myopia.
Methods
Eighty patients with myopia (80 eyes) scheduled for refractive surgery at the Ophthalmology Department of the Third People’s Hospital of Dalian from December 2024 to June 2025 were included. One eye was randomly selected for each participant, and measurements were taken using Sirius Plus and Pentacam AXL. Fourteen parameters related to the anterior segment were extracted.
Results
The differences in the measurements for anterior flat keratometry, anterior vertical/horizontal astigmatism, and anterior oblique astigmatism at 45° between the two devices were not significant, whereas those for the other parameters were significant. The intraclass correlation coefficient values for anterior flat keratometry, anterior steep keratometry, and white-to-white were 0.965, 0.963, and 0.932, respectively. The goodness-of-fit analysis showed that the R 2 values for the anterior flat keratometry, anterior steep keratometry, central corneal thickness, white-to-white, corneal volume, and anterior chamber depth were all >0.85, which demonstrated a strong correlation.
Discussion
Anterior corneal keratometry and white-to-white had good correlation and consistency, indicating they are suitable mutual reference points for clinical practice.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Ophthalmology, The Third People's Hospital of Dalian, Dalian, China.
- Department of Ophthalmology, Dalian Third People's Hospital Affiliated to Dalian University of Technology, Dalian, China.
Zitieren
APA
Ning, J., & Zhang, L. (2026). Comparison of anterior segment parameter measurements between Sirius Plus and Pentacam AXL. Frontiers in Medicine, 13. https://doi.org/10.3389/fmed.2026.1819141
Vancouver
Ning J, Zhang L. Comparison of anterior segment parameter measurements between Sirius Plus and Pentacam AXL. Frontiers in Medicine. 2026;13. doi:10.3389/fmed.2026.1819141
BibTeX
@article{Ning2026,
author = {Ning, Jiliang and Zhang, Lijun},
title = {Comparison of anterior segment parameter measurements between Sirius Plus and Pentacam AXL},
journal = {Frontiers in Medicine},
year = {2026},
volume = {13},
doi = {10.3389/fmed.2026.1819141}
}
EUVIMED