Surgical outcomes of phacoemulsification with different fluidics systems (centurion with active sentry vs. centurion gravity) in cataract patients with eye axial length above 26 mm
Frontiers in Medicine · 12.06.2025 · 12
DOI: 10.3389/fmed.2025.1554832 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Medicine, online erschienen am 12.06.2025. Verfasst von Yinan Liu und 1 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Peking University Third Hospital, Beijing, China. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Alcon.
Abstract (Originalsprache)
Introduction
During cataract phacoemulsification surgery, the Alcon Centurion with Active Sentry can achieve a more stable anterior chamber, which allows a lower intraocular pressure (IOP) setting than Centurion under gravity mode. In this randomized controlled trial, we compared these two systems’ impact on high myopia patients’ macular blood perfusion under different IOP settings. To evaluate intra-op and post-op clinical performance of the Alcon Centurion with Active Sentry and under gravity mode in cataract patients with eye axial lengths > 26 mm.
Methods
Fifty-two eyes of 43 cataract patients with axial lengths > 26 mm were enrolled and randomly divided into the active fluidics system (AFS) group using a Centurion Active Sentry handpiece under 30 mmHg IOP setting (26 eyes) and the gravity fluidics system (GFS) group in gravity mode under 80 cmH 2 O IOP setting (26 eyes). Intraoperative parameters, visual acuity, parafoveal macular vessel density and macular thickness were analyzed.
Results
We observed no significant differences in best corrected visual acuity (BCVA), macular thickness, cumulative dissipated energy (CDE), total case time, as well as ophthalmic viscoelastic devices (OVDs) usage between the two groups. However, we found less pain complaints during surgeries and a more stable macular vessel density post-surgery in the AFS group than in the GFS group, implying higher intraoperative comfort levels and less retinal ischemia during cataract surgery using Centurion with Active Sentry under low IOP settings.
Conclusion
Owing to a lower IOP setting, Centurion Vision System with Active Sentry handpiece causes less retinal ischemia and pain perception during phacoemulsification for high myopia patients. Clinical trial registration https://www.chictr.org.cn , identifier ChiCTR2400080875.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Ophthalmology, Peking University Third Hospital, Beijing, China.
Förderung
- Alcon
Zitieren
APA
Liu, Y., & Chen, X. (2025). Surgical outcomes of phacoemulsification with different fluidics systems (centurion with active sentry vs. centurion gravity) in cataract patients with eye axial length above 26 mm. Frontiers in Medicine, 12. https://doi.org/10.3389/fmed.2025.1554832
Vancouver
Liu Y, Chen X. Surgical outcomes of phacoemulsification with different fluidics systems (centurion with active sentry vs. centurion gravity) in cataract patients with eye axial length above 26 mm. Frontiers in Medicine. 2025;12. doi:10.3389/fmed.2025.1554832
BibTeX
@article{Liu2025,
author = {Liu, Yinan and Chen, Xiaoyong},
title = {Surgical outcomes of phacoemulsification with different fluidics systems (centurion with active sentry vs. centurion gravity) in cataract patients with eye axial length above 26 mm},
journal = {Frontiers in Medicine},
year = {2025},
volume = {12},
doi = {10.3389/fmed.2025.1554832}
}
EUVIMED