Diagnostic accuracy of a non-invasive spot-check hemoglobin meter, Masimo Rad-67® pulse CO-Oximeter®, in detection of anemia in antenatal care settings in Kenya
Frontiers in Global Women's Health · 14.10.2024 · 5
DOI: 10.3389/fgwh.2024.1427261 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Global Women's Health, online erschienen am 14.10.2024. Verfasst von Angela Koech und 13 weiteren Autor:innen. Beteiligte Einrichtung: Centre of Excellence in Women and Child Health, Aga Khan University, Nairobi, Kenya. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Medical Research Council, FIC NIH HHS.
Abstract (Originalsprache)
Background
Point of care hemoglobin meters play key roles in increasing access to anemia screening in antenatal care especially in settings with limited access to laboratories. We aimed to determine the diagnostic accuracy of a non-invasive spot-check hemoglobin (SpHb) meter, Masimo Rad-67® Pulse CO-Oximeter®, in the diagnosis of anemia in pregnant women attending antenatal care clinics in Kilifi, Kenya.
Methods
This was a diagnostic accuracy study that retrospectively evaluated SpHb against a validated reference standard of laboratory assessed hemoglobin (Lab Hb) by a SYSMEX XN-330 automated hematology analyzer. The study was nested within a prospective pregnancy cohort study that recruited unselected pregnant women from antenatal care clinics in two public hospitals in Kilifi County, coastal Kenya. Records with both SpHb and Lab Hb were selected from pregnancy visits between May 2021 and December 2022. Linear regression and Bland-Altman analysis were performed to compare the two tests and diagnostic accuracy parameters obtained for the diagnosis of anemia.
Results
A total of 2,975 records (from 2,203 unique participants), with paired SpHb and Lab Hb were analyzed. Linear regression showed a significant but weak positive correlation, a proportional bias of 0.44 (95% CI 0.41–0.47) and a constant of 7.59 (95% CI 7.30–7.87, p < 0.001). The median bias was 1.70 g/dl, with limits of agreement of −0.80 to 4.20. SpHb tended to be higher than Lab Hb on the low hemoglobin range but lower than Lab Hb on the high hemoglobin range. The sensitivity of SpHb in detecting anemia was 18.66%. Prevalence, specificity, positive predictive value, and negative predictive values were 46.37%, 96.77%, 83.33%, and 57.92% respectively.
Conclusion
Overall, SpHb by Masimo Rad-67® Pulse CO-Oximeter® did not accurately identify pregnant women with anemia and many cases would be missed. We would not recommend its use in antenatal care settings.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Centre of Excellence in Women and Child Health, Aga Khan University, Nairobi, Kenya.
- Faculty of Medicine and Health Sciences, Ghent University, Ghent, Belgium.
- Department of Obstetrics and Gynecology, Aga Khan University, Nairobi, Kenya.
- Department of Women and Children's Health, Kings College London, London, United Kingdom.
- Department of International Public Health, Liverpool School of Tropical Medicine (LSTM), Liverpool, United Kingdom.
- Centre for Sexual Health and HIV/AIDS Research (CeSHHAR), Harare, Zimbabwe.
Förderung
- Medical Research Council
- FIC NIH HHS
Zitieren
APA
Koech, A., Mwaniki, I., Mutunga, J., Mukhanya, M., Mwadime, E., Ochieng, M., Mwashigadi, G., Mistry, H. D., Craik, R., von Dadelszen, P., Temmerman, M., Luchters, S., Omuse, G., & PRECISE Network (2024). Diagnostic accuracy of a non-invasive spot-check hemoglobin meter, Masimo Rad-67® pulse CO-Oximeter®, in detection of anemia in antenatal care settings in Kenya. Frontiers in Global Women's Health, 5. https://doi.org/10.3389/fgwh.2024.1427261
Vancouver
Koech A, Mwaniki I, Mutunga J, Mukhanya M, Mwadime E, Ochieng M, et al. Diagnostic accuracy of a non-invasive spot-check hemoglobin meter, Masimo Rad-67® pulse CO-Oximeter®, in detection of anemia in antenatal care settings in Kenya. Frontiers in Global Women's Health. 2024;5. doi:10.3389/fgwh.2024.1427261
BibTeX
@article{Koech2024,
author = {Koech, Angela and Mwaniki, Isaac and Mutunga, Joseph and Mukhanya, Moses and Mwadime, Emily and Ochieng, Marvine and Mwashigadi, Grace and Mistry, Hiten D and Craik, Rachel and von Dadelszen, Peter and Temmerman, Marleen and Luchters, Stanley and Omuse, Geoffrey and PRECISE Network, PRECISE Network},
title = {Diagnostic accuracy of a non-invasive spot-check hemoglobin meter, Masimo Rad-67® pulse CO-Oximeter®, in detection of anemia in antenatal care settings in Kenya},
journal = {Frontiers in Global Women's Health},
year = {2024},
volume = {5},
doi = {10.3389/fgwh.2024.1427261}
}
EUVIMED