Reliable ammonia nitrogen quantification in iron- and manganese-rich groundwater: implications for groundwater quality monitoring in hydrogeological systems
Frontiers in Environmental Science · 23.06.2026 · 14
DOI: 10.3389/fenvs.2026.1842362 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Environmental Science, online erschienen am 23.06.2026. Verfasst von Zhongling Zhang und 5 weiteren Autor:innen. Die Arbeit nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Accurate determination of ammonia nitrogen (NH 3 -N) in Fe- and Mn-rich groundwater remains difficult because dissolved metals, color, turbidity, and sulfide may interfere with conventional colorimetric methods. In this study, an automated steam distillation–titration method using a Hanon K1100 fully automatic Kjeldahl nitrogen analyzer was developed and validated for NH 3 -N determination in high-interference groundwater. Method performance was evaluated using certified NH 3 -N standards, sample-volume optimization, physical and chemical interference tests, and comparative analysis of real groundwater samples from the Sanjiang Plain, China, a reduced alluvial aquifer system characterized by elevated iron and manganese. The method showed good precision and accuracy over 0.5–10.0 mg/L, with RSD values below 7% and mean recoveries of 99%–108% in the standard-solution validation. Color up to 30° and turbidity up to 182 NTU showed no significant effect on recovery. Under a 0.5 mg/L NH 3 -N background, sulfide up to 2.0 mg/L was tolerated, whereas marked negative interference appeared only at elevated Mn 2+ and Fe 2+ concentrations, with practical tolerance limits of approximately 50 mg/L and 100 mg/L, respectively. For seven representative groundwater samples, the automated Kjeldahl method produced results close to those obtained by the Nessler’s reagent colorimetric method; however, this comparison was treated as preliminary real-sample validation rather than proof of full method equivalence. The method avoids mercury-containing reagents and reduces matrix-dependent pretreatment, providing a practical and environmentally safer alternative for routine NH 3 -N monitoring in Fe- and Mn-rich groundwater.
Zitieren
APA
Zhang, Z., Wang, J., Feng, Q., Liu, T., Wang, X., & Shang, J. (2026). Reliable ammonia nitrogen quantification in iron- and manganese-rich groundwater: implications for groundwater quality monitoring in hydrogeological systems. Frontiers in Environmental Science, 14. https://doi.org/10.3389/fenvs.2026.1842362
Vancouver
Zhang Z, Wang J, Feng Q, Liu T, Wang X, Shang J. Reliable ammonia nitrogen quantification in iron- and manganese-rich groundwater: implications for groundwater quality monitoring in hydrogeological systems. Frontiers in Environmental Science. 2026;14. doi:10.3389/fenvs.2026.1842362
BibTeX
@article{Zhang2026,
author = {Zhang, Zhongling and Wang, Jiangang and Feng, Qicai and Liu, Tao and Wang, Xu and Shang, Jilong},
title = {Reliable ammonia nitrogen quantification in iron- and manganese-rich groundwater: implications for groundwater quality monitoring in hydrogeological systems},
journal = {Frontiers in Environmental Science},
year = {2026},
volume = {14},
doi = {10.3389/fenvs.2026.1842362}
}