Recurrent coronary artery in-stent restenosis caused by metal allergy: a case report
Frontiers in Cardiovascular Medicine · 09.01.2026 · 12
DOI: 10.3389/fcvm.2025.1693234 ↗Kurz gefasst
Fallbericht in der Zeitschrift Frontiers in Cardiovascular Medicine, online erschienen am 09.01.2026. Verfasst von Yang Li und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Zhongda Hospital, School of Medicine, Southeast University, Nanjing, China. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
This article describes the case of a middle-aged woman who developed recurrent in-stent restenosis (ISR) due to metal allergy. The first ISR occurred after implantation of a nickel-rich Medtronic Endeavor Resolute stent (35% nickel, 35% cobalt, and 20% chromium). A skin patch test revealed a strong nickel allergy (++), with no allergic reaction to cobalt. Nickel allergy was considered to be the cause of the first ISR. Then, a drug-coated balloon platinum–chromium alloy stent was implanted, and ISR occurred again during follow-up. The article discusses this special case and clinically relevant management options. This case highlights a scenario with refractory in-stent restenosis persisting despite various treatment strategies, including stents with different metal components, drug-coated balloons, and intensive drug therapy.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Zhongda Hospital, School of Medicine, Southeast University, Nanjing, China.
- Department of Dermatology, Zhongda Hospital, School of Medicine, Southeast University, Nanjing, China.
Zitieren
APA
Li, Y., Dong, Z., Dai, Q., Feng, Y., Chen, L., & Ma, G. (2026). Recurrent coronary artery in-stent restenosis caused by metal allergy: a case report. Frontiers in Cardiovascular Medicine, 12. https://doi.org/10.3389/fcvm.2025.1693234
Vancouver
Li Y, Dong Z, Dai Q, Feng Y, Chen L, Ma G. Recurrent coronary artery in-stent restenosis caused by metal allergy: a case report. Frontiers in Cardiovascular Medicine. 2026;12. doi:10.3389/fcvm.2025.1693234
BibTeX
@article{Li2026,
author = {Li, Yang and Dong, Zhengbang and Dai, Qiming and Feng, Yi and Chen, Lijuan and Ma, Genshan},
title = {Recurrent coronary artery in-stent restenosis caused by metal allergy: a case report},
journal = {Frontiers in Cardiovascular Medicine},
year = {2026},
volume = {12},
doi = {10.3389/fcvm.2025.1693234}
}