EUVIMEDEUVIMED entdecken
ValidierungsstudieOpen Access · CC BY 4.0

A dual-phase strategy to integrate the VITEK MS RUO database into routine clinical practice: from validation of analytical performance to implementation of an automated workflow

Hua Nian, Fushun Li, Xingqi Wang, Xiaoou Yu, Jinna Dai, Yunzuo Chu

Frontiers in Cellular and Infection Microbiology · 03.07.2026 · 16

DOI: 10.3389/fcimb.2026.1822779 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Frontiers in Cellular and Infection Microbiology, online erschienen am 03.07.2026. Verfasst von Hua Nian und 5 weiteren Autor:innen. Beteiligte Einrichtung: National Clinical Research Center for Medical Auxiliary Technology(Laboratory Medicine), Department of Laboratory Medicine, The First Hospital of Chi…. Hauptthemen laut MEDLINE: Workflow, Databases, Factual, Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization, Bacteria. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) is pivotal in clinical microbiology. The VITEK MS Research Use Only (RUO) database offers broader species coverage, yet its clinical adoption is hindered by insufficient performance validation against the approved in vitro diagnostic (IVD) database and inefficient manual operational workflows.

Objective

This two-phase study first aimed to evaluate the concordance and supplementary performance of the RUO database compared to the IVD database, and subsequently to develop and validate an automated integrated workflow to enhance laboratory efficiency.

Methods

In Phase 1 (Mar 2021–Jun 2022), parallel manual testing of 2,432 isolates was performed with both databases to analyze concordance. Based on these results, an in-house automated relay software was developed to parse IVD results and fully automate RUO reanalysis. Phase 2 (Jul–Nov 2022) prospectively incorporated 3,954 isolates to implement and assess the “IVD screening – automated RUO reanalysis” workflow.

Results

Phase 1 demonstrated high RUO-IVD concordance (95.7% species/genus agreement). The RUO database correctly identified 98.9% of isolates and provided valid supplementary identification for 84.4% (108/128) of IVD-failed cases, with Tier 2 contributing 28.9%. Phase 2 revealed that the integrated workflow increased the overall identification rate from 95.5% to 98.7%, with Tier 2 contributing an additional 14.5%. The automated software reduced reanalysis turnaround time by > 75%, saving consumables and labor.

Conclusion

The VITEK MS RUO database is a reliable and complementary tool to the IVD database. Integration with automated software creates an efficient, compliant clinical workflow, providing a practical model to enhance pathogen identification for infectious disease management. This two-phase study first sought to evaluate the concordance and supplemental identification capacity of two databases (Phase 1). Based on these performance results, an automated integrated workflow was then developed, implemented, and validated for its diagnostic efficacy and laboratory efficiency in a clinical setting (Phase 2).

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

WorkflowDatabases, FactualSpectrometry, Mass, Matrix-Assisted Laser Desorption-IonizationBacteriaSoftwareHumansAutomation, Laboratory

Stichwörter der Autor:innen

MALDI-TOF MSRUO databaseVitek MSmicrobial identificationworkflow optimization

Autor:innen und Institutionen

  • National Clinical Research Center for Medical Auxiliary Technology(Laboratory Medicine), Department of Laboratory Medicine, The First Hospital of China Medical University, Shenyang, China.

Zitieren

APA

Nian, H., Li, F., Wang, X., Yu, X., Dai, J., & Chu, Y. (2026). A dual-phase strategy to integrate the VITEK MS RUO database into routine clinical practice: from validation of analytical performance to implementation of an automated workflow. Frontiers in Cellular and Infection Microbiology, 16. https://doi.org/10.3389/fcimb.2026.1822779

Vancouver

Nian H, Li F, Wang X, Yu X, Dai J, Chu Y. A dual-phase strategy to integrate the VITEK MS RUO database into routine clinical practice: from validation of analytical performance to implementation of an automated workflow. Frontiers in Cellular and Infection Microbiology. 2026;16. doi:10.3389/fcimb.2026.1822779

BibTeX
@article{Nian2026,
  author = {Nian, Hua and Li, Fushun and Wang, Xingqi and Yu, Xiaoou and Dai, Jinna and Chu, Yunzuo},
  title = {A dual-phase strategy to integrate the VITEK MS RUO database into routine clinical practice: from validation of analytical performance to implementation of an automated workflow},
  journal = {Frontiers in Cellular and Infection Microbiology},
  year = {2026},
  volume = {16},
  doi = {10.3389/fcimb.2026.1822779}
}
Kostenlos starten →