A nested PCR-based point-of-care multiplex test for detection of bacterial pathogens in cerebrospinal fluids

Jianing Wu, Zijin Zhao, Pai Pang, Yanqing Tie, Shihao Jiao, Duoxiao Zhang, Xingyu Liu, Jie Wang, Yuxin Wang, He Tan, Mengchuan Zhao, Pei Zhao und 6 weitere

Frontiers in Cellular and Infection Microbiology · 14.04.2026 · 16

DOI: 10.3389/fcimb.2026.1781397 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Frontiers in Cellular and Infection Microbiology, online erschienen am 14.04.2026. Verfasst von Jianing Wu und 17 weiteren Autor:innen. Beteiligte Einrichtung: Graduate School, Hebei North University, Zhangjiakou, Hebei, China. Hauptthemen laut MEDLINE: Streptococcus agalactiae, Streptococcus pyogenes, Cerebrospinal Fluid, Streptococcus pneumoniae. Die Arbeit nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

To integrate a multiplex one-tube nested real-time polymerase chain reaction (mOTN-PCR) with a point-of-care testing (POCT) instrument for rapid, sensitive detection of Streptococcus pneumoniae (SPN), Streptococcus agalactiae (GBS), and Streptococcus pyogenes (GAS) in cerebrospinal fluid (CSF), achieving an automated “sample-in, result-out” workflow.

Method

The sensitivity of the mOTN-PCR was assessed using recombinant plasmids, and its specificity was evaluated using common pathogens associated with CSF infections. Simulated CSF samples were prepared using both manual nucleic acid extraction and fully automated nucleic acid extraction (POCT instrument), and then detected in parallel by mOTN-PCR and conventional real-time PCR. The clinical performance of the mOTN-PCR-POCT assay was further evaluated using 109 clinical samples (36 clinical CSF samples and 73 non-CSF clinical samples), and compared with the manual nucleic acid extraction followed by the conventional real-time PCR method.

Results

The mOTN-PCR achieved detection limits (LOD) of 5, 10, and 10 copies/μL for SPN, GBS, and GAS plasmids, respectively, with no cross-reactivity. In simulated CSF samples, the LODs were 20, 10, and 20 CFU/mL for SPN, GBS, and GAS, respectively, using manual extraction, within 4 hours. The mOTN-PCR-POCT method consistently detected all three pathogens at 20 CFU/mL, reducing total testing time to approximately 2 hours. Clinical sample testing revealed that the mOTN-PCR-POCT identified three additional positives in samples previously deemed negative by conventional real-time PCR, demonstrating superior detection capability and a more streamlined operational workflow of the mOTN-PCR-POCT.

Conclusion

The mOTN-PCR-POCT was demonstrated to be simple to operate, requires no culture, and enables rapid, highly sensitive detection of SPN, GBS, and GAS in CSF.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Streptococcus agalactiaeStreptococcus pyogenesCerebrospinal FluidStreptococcus pneumoniaeMultiplex Polymerase Chain ReactionReal-Time Polymerase Chain ReactionPoint-of-Care SystemsPoint-of-Care TestingStreptococcal InfectionsPolymerase Chain ReactionHumansSensitivity and SpecificityRapid Diagnostic TestsDNA, Bacterial

Stichwörter der Autor:innen

Streptococcus agalactiaeStreptococcus pneumoniaeStreptococcus pyogenescerebrospinal fluidmultiplex detectionmultiplex one-tube nested real-time polymerase chain reaction (mOTN-PCR)point-of-care testingpoint-of-caretesting

Substanzen

DNA, Bacterial

Autor:innen und Institutionen

  • Graduate School, Hebei North University, Zhangjiakou, Hebei, China.
  • Department of Clinical Laboratory, Hebei General Hospital, Shijiazhuang, Hebei, China.
  • National Key Laboratory of Intelligent Tracking and Forecasting for Infectious Diseases, NHC Key Laboratory of Medical Virology and Viral Diseases, National Institute for Viral Disease Control and Prevention, Chinese Ce…
  • Hebei Key Laboratory of Molecular Medicine, Shijiazhuang, Hebei, China.
  • Hebei Clinical Research Center for Laboratory Medicine, Shijiazhuang, Hebei, China.
  • Graduate School, North China University of Science and Technology, Tangshan, Hebei, China.

Zitieren

APA

Wu, J., Zhao, Z., Pang, P., Tie, Y., Jiao, S., Zhang, D., Liu, X., Wang, J., Wang, Y., Tan, H., Zhao, M., Zhao, P., Meng, J., Han, Z., Gao, S., Shen, X., Ma, X., & Feng, Z. (2026). A nested PCR-based point-of-care multiplex test for detection of bacterial pathogens in cerebrospinal fluids. Frontiers in Cellular and Infection Microbiology, 16. https://doi.org/10.3389/fcimb.2026.1781397

Vancouver

Wu J, Zhao Z, Pang P, Tie Y, Jiao S, Zhang D, et al. A nested PCR-based point-of-care multiplex test for detection of bacterial pathogens in cerebrospinal fluids. Frontiers in Cellular and Infection Microbiology. 2026;16. doi:10.3389/fcimb.2026.1781397

BibTeX
@article{Wu2026,
  author = {Wu, Jianing and Zhao, Zijin and Pang, Pai and Tie, Yanqing and Jiao, Shihao and Zhang, Duoxiao and Liu, Xingyu and Wang, Jie and Wang, Yuxin and Tan, He and Zhao, Mengchuan and Zhao, Pei and Meng, Juntao and Han, Zhiqiang and Gao, Shijue and Shen, Xinxin and Ma, Xuejun and Feng, Zhishan},
  title = {A nested PCR-based point-of-care multiplex test for detection of bacterial pathogens in cerebrospinal fluids},
  journal = {Frontiers in Cellular and Infection Microbiology},
  year = {2026},
  volume = {16},
  doi = {10.3389/fcimb.2026.1781397}
}
Kostenlos starten →