Dilute and shoot approach for toxicology testing

Kenichi Tamama

Frontiers in Chemistry · 11.12.2023 · 11

DOI: 10.3389/fchem.2023.1278313 ↗

Kurz gefasst

Übersichtsarbeit in der Zeitschrift Frontiers in Chemistry, online erschienen am 11.12.2023. Verfasst von Kenichi Tamama. Beteiligte Einrichtung: Clinical Laboratories, University of Pittsburgh Medical Center, Pittsburgh, PA, United States. Die Arbeit wurde bisher 18-mal zitiert (Crossref) und nennt 52 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Toxicology testing is performed in clinical settings, forensic settings, and for controlling doping. Drug screening is a toxicology test to determine if drugs are present in biological samples. The most common specimen type for drug testing is urine, as drugs and/or their metabolites are often more concentrated in the urine, extending the detection window of drugs. The dilute-and-shoot method is a simple procedure used in toxicology testing, where a sample is diluted before being directly injected into the liquid chromatography-mass spectrometry (LC-MS) system. This method is easy, quick, and cost-saving, and can be used for protein-poor liquid specimens such as urine. Thus, it is reasonable and attractive for busy toxicology laboratories to combine the dilute-and-shoot method with high-resolution hyphenated-MS for urine drug screening. This method has several disadvantages, including a suboptimal detection capability for certain analytes, as well as interference from co-eluting matrix components called matrix effects, in which co-eluting matrix molecules alter the ionization efficiency of the analyte molecules at the ionization source in LC-MS, altering (mostly reducing) the analyte detection capability. The matrix effect testing is essential for the validation of LC-MS-based assays. A reasonable approach to addressing these undesirable effects would be to minimize these components. The most straightforward approach is to reduce the amounts of matrix components by using a higher dilution of the specimen and a lower volume for specimen injection. Optimization of the chromatographic separation is another reasonable approach for reducing co-eluting matrix components with the analyte.

Schlagwörter

Stichwörter der Autor:innen

dilute-and-shootliquid chromatography-mass spectrometrymatrix effectsample preparationtoxicology testingurine

Autor:innen und Institutionen

  • Clinical Laboratories, University of Pittsburgh Medical Center, Pittsburgh, PA, United States.
  • Department of Pathology, University of Pittsburgh School of Medicine, Pittsburgh, PA, United States.
  • McGowan Institute for Regenerative Medicine, University of Pittsburgh, Pittsburgh, PA, United States.

Zitieren

APA

Tamama, K. (2023). Dilute and shoot approach for toxicology testing. Frontiers in Chemistry, 11. https://doi.org/10.3389/fchem.2023.1278313

Vancouver

Tamama K. Dilute and shoot approach for toxicology testing. Frontiers in Chemistry. 2023;11. doi:10.3389/fchem.2023.1278313

BibTeX
@article{Tamama2023,
  author = {Tamama, Kenichi},
  title = {Dilute and shoot approach for toxicology testing},
  journal = {Frontiers in Chemistry},
  year = {2023},
  volume = {11},
  doi = {10.3389/fchem.2023.1278313}
}
Kostenlos starten →