System accuracy evaluation of the new blood glucose monitoring meter “GLUCOCARD S onyx” beyond ISO 15197:2013/EN ISO 15197:2015 requirements and with new patient safety features

Systemgenauigkeitsbewertung des neuen Blutzuckermessgeräts „GLUCOCARD S onyx“ über die Anforderungen von ISO 15197:2013/EN ISO 15197:2015 hinaus und mit neuen Patientensicherheitsfunktionen

Daisuke Azuma, Hisashi Okuda, Beate Saeger

Frontiers in Clinical Diabetes and Healthcare · 03.04.2025 · 6

DOI: 10.3389/fcdhc.2025.1465732 ↗

Auf einen Blick

Das neue Blutzuckermessgerät GLUCOCARD S onyx zeigte in einer Evaluationsstudie eine hohe Messgenauigkeit. 99,7 % der Ergebnisse lagen innerhalb der ISO-Kriterien, 100 % in Clark Error Grid Zone A.

Fragestellung

Wie genau misst GLUCOCARD S onyx im Vergleich zu einem Referenzanalysator und erfüllt es die ISO-15197-Kriterien?

Methode

100 Fingerstich-Vollblutproben von Menschen mit Diabetes und 32 ohne Diabetes wurden mit GLUCOCARD S onyx gemessen und mit YSI2300 STAT PLUS verglichen. Die Auswertung erfolgte nach ISO 15197:2013, Abschnitt 6.3.

Ergebnisse

99,7 % (598/600) der Ergebnisse lagen innerhalb von ±15 % oder ±0,8 mmol/L (±15 mg/dL) gegenüber YSI2300 STAT PLUS. 100 % (600/600) lagen in Clark Error Grid Zone A. MARD: 4,15 %; 5,05 % bei <5,6 mmol/L und 3,65 % bei ≥5,6 mmol/L.

Schlussfolgerung der Autor:innen

GLUCOCARD S onyx zeigte eine klinisch zufriedenstellende Genauigkeit und Zuverlässigkeit und übertraf die ISO-15197:2013-Kriterien.

Kernaussagen

  • 99,7 % der Messwerte erfüllten die ISO-Genauigkeitskriterien
  • 100 % der Messwerte lagen in Clark Error Grid Zone A
  • MARD insgesamt 4,15 %

Deutsche Zusammenfassung: KI-gestützt aus dem Original-Abstract erstellt; alle Zahlen wurden automatisch mit dem Original abgeglichen. Maßgeblich ist der Original-Abstract.

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Clinical Diabetes and Healthcare, online erschienen am 03.04.2025. Verfasst von Daisuke Azuma und 2 weiteren Autor:innen. Beteiligte Einrichtung: Azuma Diabetes Clinic, Hyogo, Japan. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

Auf einen BlickAbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Blood glucose monitoring meters (BGM) have not become redundant yet. The accuracy and precision of “GLUCOCARD S onyx,” a new BGM with Bluetooth function, has been evaluated and proven to exceed the actual ISO 15197:2013/EN ISO 15197:2015 guidelines besides offering features for better patient safety and telemedicine.

Methods

100 finger-prick whole blood samples from subjects with diabetes and 32 without diabetes were collected and measured with GLUCOCARD S onyx. Plasma blood glucose levels were measured using YSI2300 STAT PLUS as reference analyzer for comparison. The evaluation followed ISO 15197:2013, section 6.3 accuracy criteria. Furthermore, the MARD factor was calculated for the overall clinical important range (with n=132 samples).

Results

The performance of GLUCOCARD S onyx was evaluated according to ISO 15197:2013, revealing that 99.7% (598/600) of the results fell within ±15% or ±0.8 mmol/L (± 15 mg/dL) of difference over the total clinically relevant glucose range compared to the YSI2300 STAT PLUS. 100% (600/600) of the measurement results over the total range fell within Clark Error Grid Zone A. An overall mean absolute relative difference (MARD) factor of 4.15% was obtained; 5.05% for glucose <5.6 mmol/L (<100 mg/dL), and 3.65% for glucose ≥5.6 mmol/L (≥100 mg/dL).

Discussion

GLUCOCARD S onyx shows clinically satisfactory accuracy and reliability, even exceeding the ISO 15197:2013 criteria, for hypoglycemic cases with glucose critically low as <3.9 mmol/L (<70 mg/dL) and hyperglycemic cases with glucose ≥10.0 mmol/L (≥180 mg/dL). Healthcare organizations as well as manufacturers are aiming to offer new BGM systems that go beyond the ISO criteria and offer systems that can be consulted instead or besides CGM (Continuous Glucose Monitoring) in case of e.g. severe hypo- and/or hyperglycemic episodes. A MARD factor of 4.15% revealed an excellent system accuracy over the total clinically relevant glucose range. With additional user-friendly features, this BGM can be seen as a useful tool for efficient diabetes therapy, especially in the event of severe blood glucose fluctuations.

Schlagwörter

Stichwörter der Autor:innen

Clark error gridEN ISO 15197:2015GLUCOCARD S onyxISO 15197:2013MARDYSI 2300accuracy evaluationblood glucose meter

Autor:innen und Institutionen

  • Azuma Diabetes Clinic, Hyogo, Japan.
  • Scientific Activity Support Team, Kyoto Laboratory, ARKRAY Inc., Kyoto, Japan.
  • Scientific Affairs Division, ARKRAY Europe B.V., Amstelveen, Netherlands.

Zitieren

APA

Azuma, D., Okuda, H., & Saeger, B. (2025). System accuracy evaluation of the new blood glucose monitoring meter “GLUCOCARD S onyx” beyond ISO 15197:2013/EN ISO 15197:2015 requirements and with new patient safety features. Frontiers in Clinical Diabetes and Healthcare, 6. https://doi.org/10.3389/fcdhc.2025.1465732

Vancouver

Azuma D, Okuda H, Saeger B. System accuracy evaluation of the new blood glucose monitoring meter “GLUCOCARD S onyx” beyond ISO 15197:2013/EN ISO 15197:2015 requirements and with new patient safety features. Frontiers in Clinical Diabetes and Healthcare. 2025;6. doi:10.3389/fcdhc.2025.1465732

BibTeX