Sensor-integrated PMMA bone cement enables multimodal data acquisition in a load-bearing orthopedic knee spacer implants
Frontiers in Bioengineering and Biotechnology · 05.11.2026 · 14
DOI: 10.3389/fbioe.2026.1960069 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 05.11.2026. Verfasst von Nguyen Bach Tran und 11 weiteren Autor:innen. Die Arbeit nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
Periprosthetic joint infection (PJI) is a severe complication of total knee arthroplasty, is a leading cause of revision surgery, and is associated with significant morbidity. Two-stage exchange using antibiotic-loaded polymethylmethacrylate (PMMA) spacers remains the clinical gold standard, yet the decision to reimplant relies largely on indirect markers and clinical judgment, as no method allows continuous in situ assessment of infection resolution. Here, we report the development of a sensorized knee spacer that transforms medical-grade PMMA bone cement from a passive structural material into an implantable multimodal sensing platform by embedding fully functional sensing and communication systems.
Methods
We developed an integration strategy addressing previously limiting factors, including exothermic polymerization, mechanical load transfer, and spatial constraints, while retaining the conventional PMMA molding principle during final spacer fabrication. A miniaturized multimodal sensor unit was integrated within PMMA cement, and the functionality of the embedded camera, spectrometer, positioning, and temperature sensors following cement integration was evaluated. The mechanical behavior of the sensor-integrated spacer was assessed under adapted dynamic biomechanical loading conditions based on ISO 14879-1:2020. In vivo functionality and longitudinal multimodal data acquisition were evaluated in a porcine pilot experiment using human knee spacer dimensions.
Results
Stable integration of the miniaturized multimodal sensor unit within PMMA cement was achieved while maintaining sensor functionality and wireless data transmission despite the thermal, mechanical, and geometric constraints of bone cement integration. The functionality of the embedded camera, spectrometer, positioning, and temperature sensors was maintained following cement integration. In vivo functionality and longitudinal multimodal data acquisition were demonstrated over 148 h in the porcine pilot experiment.
Discussion
These findings demonstrate the technical feasibility of integrating active sensing systems into load-bearing bone-cement implants under the investigated experimental conditions. The SmartSpacer provides a platform for longitudinal multimodal data acquisition during the spacer interval, establishing a technical basis for future studies investigating infection-associated biomarkers.
Zitieren
APA
Tran, N. B., Capogrosso, L., Dillitzer, C., Morandell, P., Lallinger, V., Micheler, C., Werner, J., Reiser, J., Baumgartner, C., Burgkart, R., Lazic, I., & Hayden, O. (2026). Sensor-integrated PMMA bone cement enables multimodal data acquisition in a load-bearing orthopedic knee spacer implants. Frontiers in Bioengineering and Biotechnology, 14. https://doi.org/10.3389/fbioe.2026.1960069
Vancouver
Tran NB, Capogrosso L, Dillitzer C, Morandell P, Lallinger V, Micheler C, et al. Sensor-integrated PMMA bone cement enables multimodal data acquisition in a load-bearing orthopedic knee spacer implants. Frontiers in Bioengineering and Biotechnology. 2026;14. doi:10.3389/fbioe.2026.1960069
BibTeX
@article{Tran2026,
author = {Tran, Nguyen Bach and Capogrosso, Laura and Dillitzer, Christoph and Morandell, Paul and Lallinger, Vincent and Micheler, Carina and Werner, Julia and Reiser, Judith and Baumgartner, Christine and Burgkart, Rainer and Lazic, Igor and Hayden, Oliver},
title = {Sensor-integrated PMMA bone cement enables multimodal data acquisition in a load-bearing orthopedic knee spacer implants},
journal = {Frontiers in Bioengineering and Biotechnology},
year = {2026},
volume = {14},
doi = {10.3389/fbioe.2026.1960069}
}
EUVIMED