Wireless MEMS pressure sensing for intrarenal pressure monitoring during ureteroscopy: an in vitro bench-feasibility study

Haifeng Song, Bo Wang, Zheng Xu, Xing Yang, Yubao Liu, Jianxing Li

Frontiers in Bioengineering and Biotechnology · 03.11.2026 · 14

DOI: 10.3389/fbioe.2026.1950088 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 03.11.2026. Verfasst von Haifeng Song und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Urology, Beijing Tsinghua Changgung Hospital, School of Clinical Medicine, Tsinghua University, Beijing, China. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Elevated intrarenal pressure (IRP) during retrograde intrarenal surgery is linked to infectious and parenchymal complications, motivating wireless IRP-monitoring devices that do not occupy the ureteroscope working channel.

Methods

We report an in vitro bench characterization of two units of a wireless silicon piezoresistive microelectromechanical systems pressure sensor coupled to a Bluetooth Low Energy link sampled at 10 Hz. Data were recorded under manual ureteroscope advancement and gravity irrigation at 1.0, 0.75 and 0.5 m, giving 7,630 paired samples. The device supplies digitally reported pressure readings in mmHg.

Results

Inter-unit agreement was high under manual push and at 1.0 m (Pearson r ≥ 0.996), but a systematic offset emerged at 0.5 m (mean difference –2.61 mmHg, about 40% of the steady-state mean; ICC(2,1) = 0.19 on decimated samples). Peak and steady-state pressures increased monotonically with irrigation height (R 2 = 0.997 across three points, n = 1 per point). Mean steady-state readings were about 16%–26% of the nominal hydrostatic head; this ratio describes the combined sensor–bench system response rather than intrinsic sensor sensitivity. 3-SD noise estimates over 5-s pre-irrigation baselines ranged from about 0.01 to 0.74 mmHg.

Discussion

Together these findings identify a prominent inter-unit difference at lower hydrostatic head, with offset and proportional components whose magnitude varied between recordings. Reference-manometer validation, replicated multi-height testing, and response-time, temperature and biocompatibility studies form the next steps toward clinical evaluation.

Schlagwörter

Stichwörter der Autor:innen

MEMS pressure sensorbluetooth low energyintrarenal pressureretrograde intrarenal surgerywireless sensor

Autor:innen und Institutionen

  • Department of Urology, Beijing Tsinghua Changgung Hospital, School of Clinical Medicine, Tsinghua University, Beijing, China.
  • Department of Precision Instrument, Tsinghua University, Beijing, China.
  • School of Mechanical Engineering, Dalian University of Technology, Dalian, China.

Zitieren

APA

Song, H., Wang, B., Xu, Z., Yang, X., Liu, Y., & Li, J. (2026). Wireless MEMS pressure sensing for intrarenal pressure monitoring during ureteroscopy: an in vitro bench-feasibility study. Frontiers in Bioengineering and Biotechnology, 14. https://doi.org/10.3389/fbioe.2026.1950088

Vancouver

Song H, Wang B, Xu Z, Yang X, Liu Y, Li J. Wireless MEMS pressure sensing for intrarenal pressure monitoring during ureteroscopy: an in vitro bench-feasibility study. Frontiers in Bioengineering and Biotechnology. 2026;14. doi:10.3389/fbioe.2026.1950088

BibTeX
@article{Song2026,
  author = {Song, Haifeng and Wang, Bo and Xu, Zheng and Yang, Xing and Liu, Yubao and Li, Jianxing},
  title = {Wireless MEMS pressure sensing for intrarenal pressure monitoring during ureteroscopy: an in vitro bench-feasibility study},
  journal = {Frontiers in Bioengineering and Biotechnology},
  year = {2026},
  volume = {14},
  doi = {10.3389/fbioe.2026.1950088}
}
Kostenlos starten →