Mechanical behavior of functionally zoned hybrid intracranial stent integrating laser-cut and braided architectures

Lei Ma, Mingxuan Li, Maosong Chen, Dongfeng Wang, Zhenqiang Li, Gengfan Ye, Chuan Qiao, Zhongyou Li, Gen Lv, Chong Chen, Yang Wang, Daijun Shen und 1 weitere

Frontiers in Bioengineering and Biotechnology · 02.07.2026 · 14

DOI: 10.3389/fbioe.2026.1846079 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 02.07.2026. Verfasst von Lei Ma und 12 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neurosurgery, Ningbo Medical Center Lihuili Hospital, Ningbo, Zhejiang, China. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 52 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A novel functionally zoned hybrid intracranial stent is proposed to resolve the multifaceted mechanical limitations of existing designs by heterogeneously integrating a laser-cut anchoring segment with a braided section via interwoven filaments traversing precision-machined holes. Comprehensive finite element analysis and in vitro mechanical testing demonstrate that the optimal single-ended hybrid design simultaneously achieves five critical performance metrics: intermediate radial strength (2.89 N) between pure braided (2.12 N) and laser-cut (5.04 N) stents; comparable flexibility to braided stents (bending stiffness 0.00022 N/mm vs. 0.00023 N/mm); reduced foreshortening (18.1% vs. 22.2% and 22.0% for braided configurations); elimination of the end-constriction phenomenon through the laser-cut segment’s traction effect to ensure continuous wall apposition; and preserved structural integrity during 180° bending where pure laser-cut stents fractured at 90°. Acute implantation of braided stents in rabbit subclavian arteries indirectly validates the conformability and flow-diversion capability of the braided component, confirming that this functionally zoned heterogeneous integration strategy provides a mechanically comprehensive endovascular solution for complex wide-necked intracranial aneurysms.

Schlagwörter

Stichwörter der Autor:innen

braided stentfinite element analysisflow diverterlaser-cut stentmechanical properties

Autor:innen und Institutionen

  • Department of Neurosurgery, Ningbo Medical Center Lihuili Hospital, Ningbo, Zhejiang, China.
  • Sichuan Province Biomechanical Engineering Laboratory, Sichuan University, Chengdu, China.
  • Department of Mechanical Science and Engineering, Chengdu, China.
  • Department of Biomedical Engineering, Faculty of Engineering, The Hong Kong Polytechnic University, Hong Kong, Hong Kong SAR, China.
  • Research Institute for Sports and Technology, The Hong Kong Polytechnic University, Hong Kong, Hong Kong SAR, China.
  • Chengdu Neurotrans Medical Devices Co., Ltd., Chengdu, Sichuan, China.

Zitieren

APA

Ma, L., Li, M., Chen, M., Wang, D., Li, Z., Ye, G., Qiao, C., Li, Z., Lv, G., Chen, C., Wang, Y., Shen, D., & Gao, G. (2026). Mechanical behavior of functionally zoned hybrid intracranial stent integrating laser-cut and braided architectures. Frontiers in Bioengineering and Biotechnology, 14. https://doi.org/10.3389/fbioe.2026.1846079

Vancouver

Ma L, Li M, Chen M, Wang D, Li Z, Ye G, et al. Mechanical behavior of functionally zoned hybrid intracranial stent integrating laser-cut and braided architectures. Frontiers in Bioengineering and Biotechnology. 2026;14. doi:10.3389/fbioe.2026.1846079

BibTeX
@article{Ma2026,
  author = {Ma, Lei and Li, Mingxuan and Chen, Maosong and Wang, Dongfeng and Li, Zhenqiang and Ye, Gengfan and Qiao, Chuan and Li, Zhongyou and Lv, Gen and Chen, Chong and Wang, Yang and Shen, Daijun and Gao, Ge},
  title = {Mechanical behavior of functionally zoned hybrid intracranial stent integrating laser-cut and braided architectures},
  journal = {Frontiers in Bioengineering and Biotechnology},
  year = {2026},
  volume = {14},
  doi = {10.3389/fbioe.2026.1846079}
}
Kostenlos starten →