EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Performance evaluation of biological safety cabinets: a real-world analysis

Tao Song, Shaohua Yin, Zhenlin Liu, Xiaoqi Xue, Yilin Li, Yong Li, Xingshuo Zhang, Xiangyun Qiao, Chenglin Wu, Han Zhang, Rongxi Jia, Xiaoming Liu und 6 weitere

Frontiers in Bioengineering and Biotechnology · 13.05.2026 · 14

DOI: 10.3389/fbioe.2026.1801944 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 13.05.2026. Verfasst von Tao Song und 17 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Engineering, Peking University Third Hospital, Beijing, China. Die Arbeit nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Biological safety cabinets (BSCs) are essential for protecting personnel, samples, and the environment in clinical and research laboratories. Evidence on the real-world performance of Class II BSCs in China remains limited, particularly regarding differences between domestic and imported equipment.

Methods

We conducted a cross-sectional evaluation of 360 Class II BSCs in Beijing using 1,803 performance testing records from 2018 to 2023. Performance indicators, including noise, illumination, cleanliness, inflow velocity, downflow velocity, HEPA filter integrity, and airflow smoke pattern were measured according to JG 170-2005, YY 0569-2011 and SN/T 3901–2014 standards. BSCs were assessed and classified as fully qualified performance (FQP) if the indicator met the standard without equaling the threshold, marginally qualified performance (MQP) if the indicator equaled the threshold, or unqualified if the standard was not met. Qualified rates were compared by manufacturing region, hospital level, using sector, and service life using χ 2 tests and non-parametric analyses.

Results

Among 360 Class II BSCs, the median service life was 7 years (IQR 5–10). The overall qualified rate was 70.4%, with no difference between domestic and imported BSCs (71.3% vs . 69.2%; χ 2 =0.909, P =0.340). Domestic BSCs showed higher FQP rate (55.4% vs . 40.5%; χ 2 =11.625, P =0.0007) and lower MQP rate (15.8% vs . 28.7%; χ 2 =15.087, P =0.0001), with similar pattern for noise, inflow velocity, and downflow velocity. The inflow velocity, downflow velocity, and illumination showed relatively lower qualified rates, with only inflow velocity having a higher qualified rate in domestic BSCs (85.5% vs . 80.9%; χ 2 =6.845, P =0.009). Imported BSCs performed slightly better in several key indicators, including inflow velocity, downflow velocity, and HEPA filter integrity, although differences were not statistically significant.

Discussion

These findings suggest that BSC reliability reflects an interplay between manufacturing-related characteristics and in-use performance, and that the MQP helps identify BSCs operating near minimum safety margins for proactive monitoring and timely intervention.

Schlagwörter

Stichwörter der Autor:innen

biological safety cabinetcross-sectional studylaboratory safetymarginally qualified performancemedical and health institution

Autor:innen und Institutionen

  • Department of Medical Engineering, Peking University Third Hospital, Beijing, China.
  • Department of Orthopedics, Peking University Third Hospital, Beijing, China.

Zitieren

APA

Song, T., Yin, S., Liu, Z., Xue, X., Li, Y., Li, Y., Zhang, X., Qiao, X., Wu, C., Zhang, H., Jia, R., Liu, X., Sun, Z., Yao, W., Zhang, B., Xu, F., Tian, Y., & Tian, J. (2026). Performance evaluation of biological safety cabinets: a real-world analysis. Frontiers in Bioengineering and Biotechnology, 14. https://doi.org/10.3389/fbioe.2026.1801944

Vancouver

Song T, Yin S, Liu Z, Xue X, Li Y, Li Y, et al. Performance evaluation of biological safety cabinets: a real-world analysis. Frontiers in Bioengineering and Biotechnology. 2026;14. doi:10.3389/fbioe.2026.1801944

BibTeX
@article{Song2026,
  author = {Song, Tao and Yin, Shaohua and Liu, Zhenlin and Xue, Xiaoqi and Li, Yilin and Li, Yong and Zhang, Xingshuo and Qiao, Xiangyun and Wu, Chenglin and Zhang, Han and Jia, Rongxi and Liu, Xiaoming and Sun, Zhe and Yao, Wei and Zhang, Bo and Xu, Feng and Tian, Yun and Tian, Jin},
  title = {Performance evaluation of biological safety cabinets: a real-world analysis},
  journal = {Frontiers in Bioengineering and Biotechnology},
  year = {2026},
  volume = {14},
  doi = {10.3389/fbioe.2026.1801944}
}
Kostenlos starten →