“A dual-applanation physical model to improve accuracy in goldmann tonometry by accounting for corneal biomechanics”
Frontiers in Bioengineering and Biotechnology · 29.04.2026 · 14
DOI: 10.3389/fbioe.2026.1757214 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 29.04.2026. Verfasst von Gaetano Barbaro und 3 weiteren Autor:innen. Beteiligte Einrichtung: Istitute Refractive Ophthalmology Surgery Napoli, Naples, Italy. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
We developed and analytically evaluated a novel physical-mathematical model aimed at quantifying and minimizing the systematic error ( ERR ) between true intraocular pressure ( IOPT ) and that measured by Goldmann applanation tonometry ( IOPG ), by incorporating patient-specific corneal biomechanical properties.
Methods
We constructed a model based on physical force balance, simulating the applanation of the cornea as a deformable elastic shell influenced by intraocular pressure and external forces. Four distinct forces were mathematically defined and computed: F1 – intraocular pressure acting on the posterior corneal surface; F2 – elastic resistance of the corneal tissue to flattening; F3 – adhesive force due to tear film surface tension; F4 – the net force applied by the examiner and recorded by the GAT device. The model integrates key anatomical and biomechanical parameterscentral corneal thickness, mean apical curvature radius, and corneal elastic modulus—to quantify ERR as a function of individual variability. These parameters are derived from two sequential applanation measurements ( IOPG 0 and IOPG 1 ), acquired through a custom-designed dual-zone applanation prism.
Results
Simulations demonstrated that ERR increases proportionally with central corneal thickness and elastic modulus, and decreases with increasing corneal curvature radius. For standard corneal values (thickness: 0.536 mm ; curvature: 7.15 mm ; elasticity: 0.16 MPa ), the model closely approximates IOPT . Deviations from these reference values result in systematic over- or underestimation of IOPG .
Conclusion
This analytical model provides a transparent, biomechanically grounded method for correcting GAT measurements and estimating true IOP with improved precision. It highlights the critical influence of patient-specific corneal properties and lays the foundation for personalized tonometry in clinical practice.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Istitute Refractive Ophthalmology Surgery Napoli, Naples, Italy.
- Clinica Oculistica, Pellegrini Hospital, ASL Napoli 1, Naples, Italy.
- UOC di Oculistica, Ruggi d'Aragona Hospital, Salerno, Italy.
- Clinica Oculistica, Università degli Studi di Napoli Federico II, Naples, Italy.
Zitieren
APA
Barbaro, G., Caruso, C., Troisi, M., & Costagliola, C. (2026). “A dual-applanation physical model to improve accuracy in goldmann tonometry by accounting for corneal biomechanics”. Frontiers in Bioengineering and Biotechnology, 14. https://doi.org/10.3389/fbioe.2026.1757214
Vancouver
Barbaro G, Caruso C, Troisi M, Costagliola C. “A dual-applanation physical model to improve accuracy in goldmann tonometry by accounting for corneal biomechanics”. Frontiers in Bioengineering and Biotechnology. 2026;14. doi:10.3389/fbioe.2026.1757214
BibTeX
@article{Barbaro2026,
author = {Barbaro, Gaetano and Caruso, Ciro and Troisi, Mario and Costagliola, Ciro},
title = {“A dual-applanation physical model to improve accuracy in goldmann tonometry by accounting for corneal biomechanics”},
journal = {Frontiers in Bioengineering and Biotechnology},
year = {2026},
volume = {14},
doi = {10.3389/fbioe.2026.1757214}
}
EUVIMED