Simulation of FES on the forearm with muscle-specific activation resolution
Frontiers in Bioengineering and Biotechnology · 27.06.2024 · 12
DOI: 10.3389/fbioe.2024.1384617 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 27.06.2024. Verfasst von Johanna Baier und 2 weiteren Autor:innen. Beteiligte Einrichtung: Chair for Product Development, Institute for Product and Service Engineering, Ruhr-University Bochum, Bochum, Germany. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
Functional electrical stimulation (FES) is an established method of supporting neurological rehabilitation. However, particularly on the forearm, it still cannot elicit selective muscle activations that form the basis of complex hand movements. Current research approaches in the context of selective muscle activation often attempt to enable targeted stimulation by increasing the number of electrodes and combining them in electrode arrays. In order to determine the best stimulation positions and settings, manual or semi-automated algorithms are used. This approach is limited due to experimental limitations. The supportive use of simulation studies is well-established, but existing simulation models are not suitable for analyses of selective muscle activation due to missing or arbitrarily arranged innervation zones.
Methods
This study introduces a new modeling method to design a person-specific digital twin that enables the prediction of muscle activations during FES on the forearm. The designed individual model consists of three parts: an anatomically based 3D volume conductor, a muscle-specific nerve fiber arrangement in various regions of interest (ROIs), and a standard nerve model. All processes were embedded in scripts or macros to enable automated changes to the model and the simulation setup.
Results
The experimental evaluation of simulated strength–duration diagrams showed good coincidence. The relative differences of the simulated amplitudes to the mean amplitude of the four experiments were in the same range as the inter-experimental differences, with mean values between 0.005 and 0.045. Based on these results, muscle-specific activation thresholds were determined and integrated into the simulation process. With this modification, simulated force-intensity curves showed good agreement with additionally measured curves.
Discussion
The results show that the model is suitable for simulating realistic muscle-specific activations. Since complex hand movements are physiologically composed of individual, selective muscle activations, it can be assumed that the model is also suitable for simulating these movements. Therefore, this study presents a new and very promising approach for developing new applications and products in the context of the rehabilitation of sensorimotor disorders.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Chair for Product Development, Institute for Product and Service Engineering, Ruhr-University Bochum, Bochum, Germany.
Zitieren
APA
Baier, J., Selkmann, S., & Bender, B. (2024). Simulation of FES on the forearm with muscle-specific activation resolution. Frontiers in Bioengineering and Biotechnology, 12. https://doi.org/10.3389/fbioe.2024.1384617
Vancouver
Baier J, Selkmann S, Bender B. Simulation of FES on the forearm with muscle-specific activation resolution. Frontiers in Bioengineering and Biotechnology. 2024;12. doi:10.3389/fbioe.2024.1384617
BibTeX
@article{Baier2024,
author = {Baier, Johanna and Selkmann, Sascha and Bender, Beate},
title = {Simulation of FES on the forearm with muscle-specific activation resolution},
journal = {Frontiers in Bioengineering and Biotechnology},
year = {2024},
volume = {12},
doi = {10.3389/fbioe.2024.1384617}
}