EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Atomic layer deposited TiO2 nanofilm on titanium implant for reduced the release of particles

Xiangyu Zhao, Xiaoxuan Zhang, Zilan Zhou, Fanchun Meng, Ruilin Liu, Mengyuan Zhang, Yujia Hao, Qingpeng Xie, Xiaojun Sun, Bin Zhang, Xing Wang

Frontiers in Bioengineering and Biotechnology · 26.04.2024 · 12

DOI: 10.3389/fbioe.2024.1346404 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 26.04.2024. Verfasst von Xiangyu Zhao und 10 weiteren Autor:innen. Beteiligte Einrichtung: Shanxi Medical University School and Hospital of Stomatology, Taiyuan, China. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

Titanium implants are widely used in surgeries for their biocompatibility and mechanical properties. However, excessive titanium particle release can cause implant failure. This study explores Atomic Layer Deposition (ALD) to coat commercially pure titanium (Cp-Ti) with TiO 2 , aiming to improve its frictional and corrosion resistance while reducing particle release. By comparing TiO 2 films with varying ALD cycle numbers, we assess surface properties, particle release, friction, and corrosion performance, providing insights into mitigating particle release from implants.

Methods

Cp-Ti surfaces were prepared and coated with TiO 2 films of 100, 300, and 500 ALD cycles. Surface characterization involved SEM, EDX, and XRD. Friction was tested using SEM, nanoindentation, and ICP-MS. Corrosion resistance was evaluated through immersion tests and electrochemical analysis. Cytotoxicity was assessed using BMSCs.

Results

Surface characterization revealed smoother surfaces with increased ALD cycles, confirming successful TiO 2 deposition. Friction testing showed reduced friction coefficients with higher ALD cycles, supported by nanoindentation results. Corrosion resistance improved with increasing ALD cycles, as evidenced by electrochemical tests and reduced titanium release. Cytotoxicity studies showed no significant cytotoxic effects.

Conclusion

ALD-coated TiO 2 films significantly enhance frictional and corrosion resistance of titanium implants while reducing particle release. The study underscores the importance of ALD cycle numbers in optimizing film performance, offering insights for designing implants with improved properties.

Schlagwörter

Stichwörter der Autor:innen

atomic layer depositionfriction and corrosionparticles releasetitanium dioxidetitanium implant

Autor:innen und Institutionen

  • Shanxi Medical University School and Hospital of Stomatology, Taiyuan, China.
  • State Key Laboratory of Coal Conversion, Institute of Coal Chemistry, Chinese Academy of Sciences, Taiyuan, China.
  • Department of Stomatology, The First Hospital of Shanxi Medical University, Taiyuan, China.

Zitieren

APA

Zhao, X., Zhang, X., Zhou, Z., Meng, F., Liu, R., Zhang, M., Hao, Y., Xie, Q., Sun, X., Zhang, B., & Wang, X. (2024). Atomic layer deposited TiO2 nanofilm on titanium implant for reduced the release of particles. Frontiers in Bioengineering and Biotechnology, 12. https://doi.org/10.3389/fbioe.2024.1346404

Vancouver

Zhao X, Zhang X, Zhou Z, Meng F, Liu R, Zhang M, et al. Atomic layer deposited TiO2 nanofilm on titanium implant for reduced the release of particles. Frontiers in Bioengineering and Biotechnology. 2024;12. doi:10.3389/fbioe.2024.1346404

BibTeX
@article{Zhao2024,
  author = {Zhao, Xiangyu and Zhang, Xiaoxuan and Zhou, Zilan and Meng, Fanchun and Liu, Ruilin and Zhang, Mengyuan and Hao, Yujia and Xie, Qingpeng and Sun, Xiaojun and Zhang, Bin and Wang, Xing},
  title = {Atomic layer deposited TiO2 nanofilm on titanium implant for reduced the release of particles},
  journal = {Frontiers in Bioengineering and Biotechnology},
  year = {2024},
  volume = {12},
  doi = {10.3389/fbioe.2024.1346404}
}