Comparison of kinematics and joint moments calculations for lower limbs during gait using markerless and marker-based motion capture
Frontiers in Bioengineering and Biotechnology · 12.03.2024 · 12
DOI: 10.3389/fbioe.2024.1280363 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 12.03.2024. Verfasst von Tianchen Huang und 4 weiteren Autor:innen. Beteiligte Einrichtung: Sports Biomechanics Laboratory, College of Physical Education and Health, Wenzhou University, Wenzhou, China. Die Arbeit wurde bisher 25-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective
This study aimed at quantifying the difference in kinematic and joint moments calculation for lower limbs during gait utilizing a markerless motion system (TsingVA Technology, Beijing, China) in comparison to values estimated using a marker-based motion capture system (Nokov Motion Capture System, Beijing, China).
Methods
Sixteen healthy participants were recruited for the study. The kinematic data of the lower limb during walking were acquired simultaneously based on the markerless motion capture system (120 Hz) and the marker-based motion capture system (120 Hz). The ground reaction force was recorded synchronously using a force platform (1,200 Hz). The kinematic and force data were input into Visual3D for inverse dynamics calculations.
Results
The difference in the lower limb joint center position between the two systems was the least at the ankle joint in the posterior/anterior direction, with the mean absolute deviation (MAD) of 0.74 cm. The least difference in measuring lower limb angles between the two systems was found in flexion/extension movement, and the greatest difference was found in internal/external rotation movement. The coefficient of multiple correlations (CMC) of the lower limb three joint moments for both systems exceeded or equaled 0.75, except for the ad/abduction of the knee and ankle. All the Root Mean Squared Deviation (RMSD) of the lower limb joint moment are below 18 N·m.
Conclusion
The markerless motion capture system and marker-based motion capture system showed a high similarity in kinematics and inverse dynamic calculation for lower limbs during gait in the sagittal plane. However, it should be noted that there is a notable deviation in ad/abduction moments at the knee and ankle.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Sports Biomechanics Laboratory, College of Physical Education and Health, Wenzhou University, Wenzhou, China.
- Laboratory of Biomechanics and Rehabilitation Engineering, School of Medicine, Tongji University, Shanghai, China.
- TsingVA (Beijing) Technology Co., Ltd., Beijing, China.
- Key Laboratory of Exercise and Health Sciences, Ministry of Education, Shanghai University of Sport, Shanghai, China.
Zitieren
APA
Huang, T., Ruan, M., Huang, S., Fan, L., & Wu, X. (2024). Comparison of kinematics and joint moments calculations for lower limbs during gait using markerless and marker-based motion capture. Frontiers in Bioengineering and Biotechnology, 12. https://doi.org/10.3389/fbioe.2024.1280363
Vancouver
Huang T, Ruan M, Huang S, Fan L, Wu X. Comparison of kinematics and joint moments calculations for lower limbs during gait using markerless and marker-based motion capture. Frontiers in Bioengineering and Biotechnology. 2024;12. doi:10.3389/fbioe.2024.1280363
BibTeX
@article{Huang2024,
author = {Huang, Tianchen and Ruan, Mianfang and Huang, Shangjun and Fan, Linlin and Wu, Xie},
title = {Comparison of kinematics and joint moments calculations for lower limbs during gait using markerless and marker-based motion capture},
journal = {Frontiers in Bioengineering and Biotechnology},
year = {2024},
volume = {12},
doi = {10.3389/fbioe.2024.1280363}
}
EUVIMED