Biomechanical analysis and clinical observation of 3D-printed acetabular prosthesis for the acetabular reconstruction of total hip arthroplasty in Crowe III hip dysplasia

Yuchen Liu, Fuyang Wang, Jiawei Ying, Minghao Xu, Yuan Wei, Junlei Li, Hui Xie, Dewei Zhao, Liangliang Cheng

Frontiers in Bioengineering and Biotechnology · 15.09.2023 · 11

DOI: 10.3389/fbioe.2023.1219745 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 15.09.2023. Verfasst von Yuchen Liu und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedics, Affiliated Zhongshan Hospital of Dalian University, Dalian, China. Die Arbeit wurde bisher 18-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

This study aimed to evaluate the biomechanical effectiveness of 3D-printed integrated acetabular prosthesis (IAP) and modular acetabular prosthesis (MAP) in reconstructing the acetabulum for patients with Crowe III developmental dysplasia of the hip (DDH). The results of this study can provide a theoretical foundation for the treatment of Crowe III DDH in total hip arthroplasty (THA).

Methods

Finite element (FE) analysis models were created to reconstruct Crowe III DDH acetabular defects using IAP and MAP. The contact stress and relative micromotion between the acetabular prosthesis and the host bone were analyzed by gradually loading in three increments (210 N, 2100 N, and 4200 N). In addition, five patients with Crowe III DDH who underwent IAP acetabular reconstruction were observed.

Results

At the same load, the peak values of IAP contact stress and relative micromotion were lower than those of MAP acetabular reconstruction. Under jogging load, the MAP metal augment’s peak stress exceeded porous tantalum yield strength, and the risk of prosthesis fracture was higher. The peak stress in the bone interface in contact with the MAP during walking and jogging was higher than that in the cancellous bone, while that of IAP was higher than that of the cancellous bone only under jogging load, so the risk of MAP cancellous bone failure was greater. Under jogging load, the relative micromotion of the MAP reconstruction acetabular implant was 45.2 μm, which was not conducive to bone growth, while under three different loads, the relative micromotion of the IAP acetabular implant was 1.5–11.2 μm, all <40 μm, which was beneficial to bone growth. Five patients with IAP acetabular reconstruction were followed up for 11.8 ± 3.4 months, and the Harris score of the last follow-up was 85.4 ± 5.5. The imaging results showed good stability of all prostheses with no adverse conditions observed.

Conclusion

Compared with acetabular reconstruction with MAP, IAP has a lower risk of loosening and fracture, as well as a better long-term stability. The application of IAP is an ideal acetabular reconstruction method for Crowe III DDH.

Schlagwörter

Stichwörter der Autor:innen

3D printingacetabular reconstructiondevelopmental dysplasia of the hipfinite element analysisintegral acetabular prosthesismodular acetabular prosthesis

Autor:innen und Institutionen

  • Department of Orthopedics, Affiliated Zhongshan Hospital of Dalian University, Dalian, China.
  • Affiliated Zhongshan Hospital of Dalian University, Dalian, China.

Zitieren

APA

Liu, Y., Wang, F., Ying, J., Xu, M., Wei, Y., Li, J., Xie, H., Zhao, D., & Cheng, L. (2023). Biomechanical analysis and clinical observation of 3D-printed acetabular prosthesis for the acetabular reconstruction of total hip arthroplasty in Crowe III hip dysplasia. Frontiers in Bioengineering and Biotechnology, 11. https://doi.org/10.3389/fbioe.2023.1219745

Vancouver

Liu Y, Wang F, Ying J, Xu M, Wei Y, Li J, et al. Biomechanical analysis and clinical observation of 3D-printed acetabular prosthesis for the acetabular reconstruction of total hip arthroplasty in Crowe III hip dysplasia. Frontiers in Bioengineering and Biotechnology. 2023;11. doi:10.3389/fbioe.2023.1219745

BibTeX
@article{Liu2023,
  author = {Liu, Yuchen and Wang, Fuyang and Ying, Jiawei and Xu, Minghao and Wei, Yuan and Li, Junlei and Xie, Hui and Zhao, Dewei and Cheng, Liangliang},
  title = {Biomechanical analysis and clinical observation of 3D-printed acetabular prosthesis for the acetabular reconstruction of total hip arthroplasty in Crowe III hip dysplasia},
  journal = {Frontiers in Bioengineering and Biotechnology},
  year = {2023},
  volume = {11},
  doi = {10.3389/fbioe.2023.1219745}
}
Kostenlos starten →