The influence of over-distraction on biomechanical response of cervical spine post anterior interbody fusion: a comprehensive finite element study
Frontiers in Bioengineering and Biotechnology · 15.08.2023 · 11
DOI: 10.3389/fbioe.2023.1217274 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 15.08.2023. Verfasst von Chih-Hsiu Cheng und 4 weiteren Autor:innen. Beteiligte Einrichtung: School of Physical Therapy and Graduate Institute of Rehabilitation Science, College of Medicine, Chang Gung University, Taoyuan, Taiwan. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 69 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
Anterior cervical discectomy and fusion (ACDF) has been considered as the gold standard surgical treatment for cervical degenerative pathologies. Some surgeons tend to use larger-sized interbody cages during ACDF to restore the index intervertebral disc height, hence, this study evaluated the effect of larger-sized interbody cages on the cervical spine with ACDF under both static and cyclic loading.
Method
Twenty pre-operative personalized poro-hyperelastic finite element (FE) models were developed. ACDF post-operative models were then constructed and four clinical scenarios (i.e., 1) No-distraction; 2) 1 mm distraction; 3) 2 mm distraction; and 4) 3 mm distraction) were predicted for each patient. The biomechanical responses at adjacent spinal levels were studied subject to static and cyclic loading. Non-parametric Friedman statistical comparative tests were performed and the p values less than 0.05 were reflected as significant.
Results
The calculated intersegmental range of motion (ROM) and intradiscal pressure (IDP) from 20 pre-operative FE models were within the overall ranges compared to the available data from literature. Under static loading, greater ROM, IDP, facet joint force (FJF) values were detected post ACDF, as compared with pre-op. Over-distraction induced significantly higher IDP and FJF in both upper and lower adjacent levels in extension. Higher annulus fibrosus stress and strain values, and increased disc height and fluid loss at the adjacent levels were observed in ACDF group which significantly increased for over-distraction groups. Discussion: it was concluded that using larger-sized interbody cages (the height of ≥2 mm of the index disc height) can result in remarkable variations in biomechanical responses of adjacent levels, which may indicate as risk factor for adjacent segment disease. The results of this comprehensive FE investigation using personalized modeling technique highlight the importance of selecting the appropriate height of interbody cage in ACDF surgery.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- School of Physical Therapy and Graduate Institute of Rehabilitation Science, College of Medicine, Chang Gung University, Taoyuan, Taiwan.
- Bone and Joint Research Center, Chang Gung Memorial Hospital, Linkou, Taiwan.
- Department of Orthopedic Surgery, Chang Gung Memorial Hospital, Linkou, Taiwan.
- Department of Biomedical Engineering, Science and Research Branch, Islamic Azad University, Tehran, Iran.
Zitieren
APA
Cheng, C. H., Chiu, P. Y., Chen, H. B., Niu, C. C., & Nikkhoo, M. (2023). The influence of over-distraction on biomechanical response of cervical spine post anterior interbody fusion: a comprehensive finite element study. Frontiers in Bioengineering and Biotechnology, 11. https://doi.org/10.3389/fbioe.2023.1217274
Vancouver
Cheng CH, Chiu PY, Chen HB, Niu CC, Nikkhoo M. The influence of over-distraction on biomechanical response of cervical spine post anterior interbody fusion: a comprehensive finite element study. Frontiers in Bioengineering and Biotechnology. 2023;11. doi:10.3389/fbioe.2023.1217274
BibTeX
@article{Cheng2023,
author = {Cheng, Chih-Hsiu and Chiu, Ping-Yeh and Chen, Hung-Bin and Niu, Chi-Chien and Nikkhoo, Mohammad},
title = {The influence of over-distraction on biomechanical response of cervical spine post anterior interbody fusion: a comprehensive finite element study},
journal = {Frontiers in Bioengineering and Biotechnology},
year = {2023},
volume = {11},
doi = {10.3389/fbioe.2023.1217274}
}