2D-3D reconstruction of the proximal femur from DXA scans: Evaluation of the 3D-Shaper software
Frontiers in Bioengineering and Biotechnology · 01.03.2023 · 11
DOI: 10.3389/fbioe.2023.1111020 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 01.03.2023. Verfasst von Alice Dudle und 5 weiteren Autor:innen. Beteiligte Einrichtung: ARTORG Center for Biomedical Engineering Research, University of Bern, Bern, Switzerland. Die Arbeit wurde bisher 35-mal zitiert (Crossref) und nennt 52 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung.
Abstract (Originalsprache)
Introduction
Osteoporosis is currently diagnosed based on areal bone mineral density (aBMD) computed from 2D DXA scans. However, aBMD is a limited surrogate for femoral strength since it does not account for 3D bone geometry and density distribution. QCT scans combined with finite element (FE) analysis can deliver improved femoral strength predictions. However, non-negligible radiation dose and high costs prevent a systematic usage of this technique for screening purposes. As an alternative, the 3D-Shaper software (3D-Shaper Medical, Spain) reconstructs the 3D shape and density distribution of the femur from 2D DXA scans. This approach could deliver a more accurate estimation of femoral strength than aBMD by using FE analysis on the reconstructed 3D DXA.
Methods
Here we present the first independent evaluation of the software, using a dataset of 77 ex vivo femora. We extend a prior evaluation by including the density distribution differences, the spatial correlation of density values and an FE analysis. Yet, cortical thickness is left out of this evaluation, since the cortex is not resolved in our FE models.
Results
We found an average surface distance of 1.16 mm between 3D DXA and QCT images, which shows a good reconstruction of the bone geometry. Although BMD values obtained from 3D DXA and QCT correlated well ( r 2 = 0.92), the 3D DXA BMD were systematically lower. The average BMD difference amounted to 64 mg/cm 3 , more than one-third of the 3D DXA BMD. Furthermore, the low correlation ( r 2 = 0.48) between density values of both images indicates a limited reconstruction of the 3D density distribution. FE results were in good agreement between QCT and 3D DXA images, with a high coefficient of determination ( r 2 = 0.88). However, this correlation was not statistically different from a direct prediction by aBMD. Moreover, we found differences in the fracture patterns between the two image types. QCT-based FE analysis resulted mostly in femoral neck fractures and 3D DXA-based FE in subcapital or pertrochanteric fractures.
Discussion
In conclusion, 3D-Shaper generates an altered BMD distribution compared to QCT but, after careful density calibration, shows an interesting potential for deriving a standardized femoral strength from a DXA scan.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- ARTORG Center for Biomedical Engineering Research, University of Bern, Bern, Switzerland.
- Division of Anatomy, Gottfried Schatz Research Center, Medical University of Graz, Graz, Austria.
- Division of Anatomy, Center for Anatomy and Cell Biology, Medical University of Vienna, Vienna, Austria.
- Division of Bone Diseases, Geneva University Hospitals (HUG), Geneva, Switzerland.
- Department of Osteoporosis, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
Förderung
- Schweizerischer Nationalfonds zur Förderung der Wissenschaftlichen Forschung
Zitieren
APA
Dudle, A., Gugler, Y., Pretterklieber, M., Ferrari, S., Lippuner, K., & Zysset, P. (2023). 2D-3D reconstruction of the proximal femur from DXA scans: Evaluation of the 3D-Shaper software. Frontiers in Bioengineering and Biotechnology, 11. https://doi.org/10.3389/fbioe.2023.1111020
Vancouver
Dudle A, Gugler Y, Pretterklieber M, Ferrari S, Lippuner K, Zysset P. 2D-3D reconstruction of the proximal femur from DXA scans: Evaluation of the 3D-Shaper software. Frontiers in Bioengineering and Biotechnology. 2023;11. doi:10.3389/fbioe.2023.1111020
BibTeX
@article{Dudle2023,
author = {Dudle, Alice and Gugler, Yvan and Pretterklieber, Michael and Ferrari, Serge and Lippuner, Kurt and Zysset, Philippe},
title = {2D-3D reconstruction of the proximal femur from DXA scans: Evaluation of the 3D-Shaper software},
journal = {Frontiers in Bioengineering and Biotechnology},
year = {2023},
volume = {11},
doi = {10.3389/fbioe.2023.1111020}
}
EUVIMED