EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

A dual-band RF exposure system for real-time electrophysiological recordings using microelectrode array device

Noemi Dolciotti, Carmen Pisano, Alessandra Paffi, Margherita Cioni, Giada Guerra, Isabella Zironi, Daniel Remondini, Giorgio Aicardi, Micaela Liberti, Francesca Apollonio

Frontiers in Antennas and Propagation · 2026

DOI: 10.3389/fanpr.2026.1832099 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Antennas and Propagation, 2026. Verfasst von Noemi Dolciotti und 9 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Fifth-generation (5G) wireless communication systems operating in the FR1 band have increased interest in studying bioelectromagnetic interactions under controlled exposure conditions. MicroElectrode Array (MEA)-based electrophysiology provides a suitable approach for real-time investigations of neuronal activity, although its integration with RF exposure systems remains technically challenging.

Methods

This work presents the design, integration, and numerical characterization of a dual-band RF exposure system for real-time electrophysiological recordings using MEA devices. The system operates at 0.7 GHz and 3.5 GHz and is based on an open-ended coaxial radiator, termed Localized Radiating-End Coaxial line (LoREC). Full-wave COMSOL simulations were performed to evaluate field confinement, dosimetric quantities (SAR), and compatibility with the MEA acquisition setup.

Results

The proposed exposure system achieved stable impedance matching and effective confinement of the electromagnetic field within the biological target region. SAR analysis showed frequency-dependent exposure patterns with reproducible and spatially consistent distributions across the active electrode area. Simulations also demonstrated negligible residual fields in proximity to the acquisition electronics, indicating limited electromagnetic interference with electrophysiological recordings.

Discussion

The proposed exposure system provides a controlled and reproducible exposure architecture suitable for real-time RF exposure and MEA electrophysiological measurements under physiological conditions. The numerical results demonstrated stable near-field confinement within the MEA culture region, predictable SAR distributions, and negligible residual fields on the acquisition electronics across the investigated frequency range. The system is therefore suitable for investigating frequency-dependent bioelectromagnetic effects associated with 5G-FR1 technologies.

Zitieren

APA

Dolciotti, N., Pisano, C., Paffi, A., Cioni, M., Guerra, G., Zironi, I., Remondini, D., Aicardi, G., Liberti, M., & Apollonio, F. (2026). A dual-band RF exposure system for real-time electrophysiological recordings using microelectrode array device. Frontiers in Antennas and Propagation. https://doi.org/10.3389/fanpr.2026.1832099

Vancouver

Dolciotti N, Pisano C, Paffi A, Cioni M, Guerra G, Zironi I, et al. A dual-band RF exposure system for real-time electrophysiological recordings using microelectrode array device. Frontiers in Antennas and Propagation. 2026. doi:10.3389/fanpr.2026.1832099

BibTeX
@article{Dolciotti2026,
  author = {Dolciotti, Noemi and Pisano, Carmen and Paffi, Alessandra and Cioni, Margherita and Guerra, Giada and Zironi, Isabella and Remondini, Daniel and Aicardi, Giorgio and Liberti, Micaela and Apollonio, Francesca},
  title = {A dual-band RF exposure system for real-time electrophysiological recordings using microelectrode array device},
  journal = {Frontiers in Antennas and Propagation},
  year = {2026},
  doi = {10.3389/fanpr.2026.1832099}
}
Kostenlos starten →