Fachartikel in der Zeitschrift Annals of Laboratory Medicine, online erschienen am 30.06.2023. Verfasst von Hyunji Kim und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, Seoul National University Bundang Hospital and Seoul National University College of Medicine, Seoul, Korea. Hauptthemen laut MEDLINE: Laboratories, Clinical Laboratory Services. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 13 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: National Research Foundation of Korea, Ministry of Science and ICT, South Korea, Asan Institute for Life Sciences, Asan Medical Center.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The adoption of high-sensitivity flow cytometry (HSFC) in routine laboratory settings has been slow owing to concerns regarding the reliability and reproducibility of results. Validation is an essential prerequisite for conducting assays, and implementing the CLSI guidelines has been confusing, primarily because many aspects are not yet established. We aimed to validate an HSFC protocol for detecting follicular helper T (Tfh) cells in a real-world laboratory environment. The analytical validity of the Tfh cell panel was ensured through rigorous testing, including evaluations of precision, stability, carryover, and sensitivity, following the CLSI H62 guidelines. We found that Tfh cells, present in very small numbers in the blood, could be sufficiently detected through HSFC, and concerns about the reliability and reproducibility of the results in real-world laboratories could be solved through systematic validation. Establishing the lower limit of quantification (LLOQ) is a critical step in HSFC evaluations. By selecting an appropriate sample, for example, collecting residual cells from CD4 isolation in our experiment and using them as low-level samples, the LLOQ could be accurately established. The strategic validation of flow cytometry panels can facilitate the adoption of HSFC in clinical laboratories, even with limited resources.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LaboratoriesClinical Laboratory ServicesHumansReproducibility of ResultsFlow Cytometry

Stichwörter der Autor:innen

CLSI H26Follicular helper T cellsHigh-sensitivity flow cytometryValidation

Autor:innen und Institutionen

  • Department of Laboratory Medicine, Seoul National University Bundang Hospital and Seoul National University College of Medicine, Seoul, Korea.
  • Department of Surgery, Division of Kidney and Pancreas Transplantation, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.
  • Department of Laboratory Medicine, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Korea.

ORCID-Profile: Hyunji Kim, Sung Shin, Sang-Hyun Hwang, Miyoung Kim, Young-Uk Cho, Seongsoo Jang

Förderung

  • National Research Foundation of Korea
  • Ministry of Science and ICT, South Korea
  • Asan Institute for Life Sciences, Asan Medical Center

Zitieren

APA

Kim, H., Shin, S., Hwang, S. H., Kim, M., Cho, Y. U., & Jang, S. (2023). Validation of High-sensitivity Flow Cytometry for Reliable Immune Cell Analysis in Real-world Laboratory Settings. Annals of Laboratory Medicine, 43(6), 620-624. https://doi.org/10.3343/alm.2023.43.6.620

Vancouver

Kim H, Shin S, Hwang SH, Kim M, Cho YU, Jang S. Validation of High-sensitivity Flow Cytometry for Reliable Immune Cell Analysis in Real-world Laboratory Settings. Annals of Laboratory Medicine. 2023;43(6):620-624. doi:10.3343/alm.2023.43.6.620

BibTeX
@article{Kim2023,
  author = {Kim, Hyunji and Shin, Sung and Hwang, Sang-Hyun and Kim, Miyoung and Cho, Young-Uk and Jang, Seongsoo},
  title = {Validation of High-sensitivity Flow Cytometry for Reliable Immune Cell Analysis in Real-world Laboratory Settings},
  journal = {Annals of Laboratory Medicine},
  year = {2023},
  volume = {43},
  number = {6},
  pages = {620--624},
  doi = {10.3343/alm.2023.43.6.620}
}
Kostenlos starten →