Comparison of the New VITEK MS PRIME System with the Matrix-Assisted Laser Desorption Ionization Biotyper Microflex LT for the Identification of Microorganisms
Annals of Laboratory Medicine · 2023
DOI: 10.3343/alm.2023.43.6.574 ↗Kurz gefasst
Fachartikel in der Zeitschrift Annals of Laboratory Medicine, 2023. Verfasst von Patrick Grohs und 6 weiteren Autor:innen. Beteiligte Einrichtung: Bacteriology Department, Microbiology Service, Assistance Publique-Hôpitaux de Paris, Hôpital Européen Georges Pompidou, Paris, France. Hauptthemen laut MEDLINE: Saccharomyces cerevisiae, Laboratories. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background
Matrix-assisted laser desorption ionization (MALDI) is the cornerstone of bacterial identification. The performance of a new MALDI time-of-flight mass spectrometry VITEK MS PRIME (VMS-P) system was compared with that of the MALDI Biotyper Microflex LT (MBT) system, which is routinely used in our laboratory.
Methods
Sixteen bacterial and yeast reference strains cultured in 20 different media were analyzed over 10 consecutive rounds using both systems. Bacterial and yeast isolates from the routine workflow were processed using both systems. Microcolonies were identified after a 4-hour agar subculture from positive blood culture bottles, without extraction.
Results
To determine the repeatability based on the reference strains, 1,190 spots were processed using each system. Correct identification was achieved for 94.0% (MBT) and 98.4% (VMS-P; P
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Bacteriology Department, Microbiology Service, Assistance Publique-Hôpitaux de Paris, Hôpital Européen Georges Pompidou, Paris, France.
- Parasite and Mycology Department, Microbiology Service, Assistance Publique-Hôpitaux de Paris, Hôpital Européen Georges Pompidou, Paris, France.
- Mycology Department, Université de Paris, France.
ORCID-Profile: Patrick Grohs, Elodie Remaud, Cybill Lath, Kim Vuong, Marie-Lize Parolini, Eric Dannaoui, Isabelle Podglajen
Zitieren
APA
Grohs, P., Remaud, E., Lath, C., Vuong, K., Parolini, M. L., Dannaoui, E., & Podglajen, I. (2023). Comparison of the New VITEK MS PRIME System with the Matrix-Assisted Laser Desorption Ionization Biotyper Microflex LT for the Identification of Microorganisms. Annals of Laboratory Medicine. https://doi.org/10.3343/alm.2023.43.6.574
Vancouver
Grohs P, Remaud E, Lath C, Vuong K, Parolini ML, Dannaoui E, et al. Comparison of the New VITEK MS PRIME System with the Matrix-Assisted Laser Desorption Ionization Biotyper Microflex LT for the Identification of Microorganisms. Annals of Laboratory Medicine. 2023. doi:10.3343/alm.2023.43.6.574
BibTeX
@article{Grohs2023,
author = {Grohs, Patrick and Remaud, Elodie and Lath, Cybill and Vuong, Kim and Parolini, Marie-Lize and Dannaoui, Eric and Podglajen, Isabelle},
title = {Comparison of the New VITEK MS PRIME System with the Matrix-Assisted Laser Desorption Ionization Biotyper Microflex LT for the Identification of Microorganisms},
journal = {Annals of Laboratory Medicine},
year = {2023},
doi = {10.3343/alm.2023.43.6.574}
}