Intraoperative Validation of Left Internal Mammary Artery Graft by Flourescence Imaging Technique
Cardiovascular Journal · 2018
DOI: 10.3329/cardio.v10i2.36282 ↗Kurz gefasst
Fachartikel in der Zeitschrift Cardiovascular Journal, 2018. Verfasst von AM Asif Rahim und 5 weiteren Autor:innen.
Abstract (Originalsprache)
Background
In coronary artery bypass surgery (CABG) left internal mammary artery (LIMA) is considered gold standard conduit of choice for myocardial revascularization. Graft failure following CABG reduces cardiac mortality and morbidity both in short and long term. Although conventional angiography is gold standard for assessing graft patency but rarely available in same operating room. So intraoperative florescence imaging could be an efficient and reliable method of assessing the patency of graft.
Methods
This study was conducted between July 2013 to June 2014 in the Department of Cardiac Surgery of National Institute of Cardiovascular Diseases (NICVD). Thirty six LIMA grafts were assessed by using IFI system .ICG administered through CVP line and imaging acquired during pass of the ICG through field of view graft flow. Quality of anastomosis was intra operatively to validate graft.
Results
Mean age of study population was 54±8.38 years. Per operative assessment of LIMA to left anterior descending artery revealed 32 (88.8%) patent anastomosis, narrowing of anastomosis was found in 2 (5.55% ) patients.
Conclusion
Intraoperative fluorescence imaging is an effective and inexpensive way to validate patency of LIMA graft.Cardiovasc. j. 2018; 10(2): 145-149
Zitieren
APA
Rahim, A. A., Pervez, M., Kibria, M., Talukder, M., Zaman, A., & Hoque, Q. (2018). Intraoperative Validation of Left Internal Mammary Artery Graft by Flourescence Imaging Technique. Cardiovascular Journal. https://doi.org/10.3329/cardio.v10i2.36282
Vancouver
Rahim AA, Pervez M, Kibria M, Talukder M, Zaman A, Hoque Q. Intraoperative Validation of Left Internal Mammary Artery Graft by Flourescence Imaging Technique. Cardiovascular Journal. 2018. doi:10.3329/cardio.v10i2.36282
BibTeX
@article{Rahim2018,
author = {Rahim, AM Asif and Pervez, MAH and Kibria, MG and Talukder, MSH and Zaman, A and Hoque, QM},
title = {Intraoperative Validation of Left Internal Mammary Artery Graft by Flourescence Imaging Technique},
journal = {Cardiovascular Journal},
year = {2018},
doi = {10.3329/cardio.v10i2.36282}
}
EUVIMED