Impact of Monte Carlo statistical uncertainty on EGSnrc-based 6 MV beam modeling of an Elekta Synergy linear accelerator
Radiation Sciences in Oncology · 27.04.2026 · 1 (1) · 23-30
DOI: 10.33066/rso.2026.2 ↗Kurz gefasst
Fachartikel in der Zeitschrift Radiation Sciences in Oncology, online erschienen am 27.04.2026. Verfasst von Esin Gündem und 1 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective
Monte Carlo (MC) simulation is widely used for accurate modeling of medical linear accelerators; however, studies quantifying the effects of statistical uncertainty on beam stability and dosimetric accuracy are limited. This study investigates the effect of varying MC statistical uncertainty levels on the spectral stability, Percentage Depth Dose (PDD), and lateral beam profile parameters of the 6 MV photon beam from the Elekta Synergy linear accelerator using the Electron–Gamma–Shower (EGS)nrc Code System. Materials and Methods: The EGSnrc-based model of the Elekta Synergy 6 MV beam was simulated using the manufacturer’s treatment head specifications. Simulations were performed over a 10×10 cm² area, with particle histories ranging from 10⁴ to 10⁹, corresponding to statistical uncertainties of 4.9% to 0.1%. The primary electron beam was defined as an initial energy of 5.8 MeV and a Gaussian spatial distribution Full Width at Half Maximum (FWHM) of 1 mm). Phase space files were used to calculate dose distributions in the virtual water phantom. Simulated PDD curves and lateral profiles were compared with measurements obtained using the motorized water phantom and CC13 ionization chambers.
Results
In our study, high statistical uncertainty (≥1–2%) led to significant stochastic fluctuations in the photon energy spectrum and the PDD curve. As uncertainty decreased below 0.5%, spectral and depth-dose stability improved, and agreement with measured dosimetric parameters (dmax, d5, d10, TPR20,10) increased. Flatness and symmetry were sensitive to statistical noise, whereas the FWHM showed limited dependence.
Conclusion
Statistical uncertainty is shown to influence the photon energy spectrum, PDD characteristics, and profile-based beam metrics. For accurate EGSnrc-based modeling and reliable parameter optimization, uncertainty levels below ~0.5% are recommended to ensure convergence, whereas values below 1% appear sufficient to achieve stable and reproducible results in high-dose regions.
Zitieren
APA
Gündem, E., & Dirican, B. (2026). Impact of Monte Carlo statistical uncertainty on EGSnrc-based 6 MV beam modeling of an Elekta Synergy linear accelerator. Radiation Sciences in Oncology, 1(1), 23-30. https://doi.org/10.33066/rso.2026.2
Vancouver
Gündem E, Dirican B. Impact of Monte Carlo statistical uncertainty on EGSnrc-based 6 MV beam modeling of an Elekta Synergy linear accelerator. Radiation Sciences in Oncology. 2026;1(1):23-30. doi:10.33066/rso.2026.2
BibTeX
@article{Gndem2026,
author = {Gündem, Esin and Dirican, Bahar},
title = {Impact of Monte Carlo statistical uncertainty on EGSnrc-based 6 MV beam modeling of an Elekta Synergy linear accelerator},
journal = {Radiation Sciences in Oncology},
year = {2026},
volume = {1},
number = {1},
pages = {23--30},
doi = {10.33066/rso.2026.2}
}