EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Design and Experimental Validation of a Mixed Signal PSoC Based System on Chip for Monitoring of Neonatal Incubator Parameters

N. N. Kumbhar, Dr. S. K. Tilekar

International Journal of Scientific Research in Science, Engineering and Technology · 22.08.2026 · 13 (4) · 335-346

DOI: 10.32628/ijsrset26134131 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Scientific Research in Science, Engineering and Technology, online erschienen am 22.08.2026. Verfasst von N. N. Kumbhar und 1 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Neonatal incubators provide a controlled microenvironment for premature and critically ill newborn infants; however, continuous monitoring of multiple environmental and physiological parameters generally requires separate sensing and monitoring units. This work presents the design and experimental validation of a mixed signal System on Chip (SoC) based on the CY8CKIT-062-WiFi-BT PSoC 62 Pioneer Kit for integrated real time monitoring of neonatal incubator parameters. The designed system combines four sensing modules for temperature, relative humidity, phototherapy light intensity and pulse rate within a single programmable embedded platform. An LM35 temperature sensor, SY-HS-220 humidity sensor, BPW34 silicon PIN photodiode and GE SpO₂ neonatal probe were interfaced with the programmable analog resources and integrated 12-bit successive approximation register analog to digital converter of the PSoC 62. Embedded firmware developed using PSoC Creator 4.4 performs sensor acquisition, calibration, signal processing, parameter computation, visualization and serial data communication. The measured parameters are displayed on a TFT graphical display and simultaneously transmitted through UART to a computer for real time monitoring and data logging. The designed system was experimentally evaluated under practical hospital operating conditions at Vighnaharta Multi Speciality Hospital, Karjat, Dixit Hospital, Phaltan and Niramay Hospital, Satara. Comparative measurements demonstrated close agreement between the designed system and the corresponding hospital reference instruments. The results establish the practical feasibility of using a programmable mixed signal SoC for compact, continuous and integrated neonatal incubator monitoring.

Zitieren

APA

Kumbhar, N. N., & Tilekar, D. S. K. (2026). Design and Experimental Validation of a Mixed Signal PSoC Based System on Chip for Monitoring of Neonatal Incubator Parameters. International Journal of Scientific Research in Science, Engineering and Technology, 13(4), 335-346. https://doi.org/10.32628/ijsrset26134131

Vancouver

Kumbhar NN, Tilekar DSK. Design and Experimental Validation of a Mixed Signal PSoC Based System on Chip for Monitoring of Neonatal Incubator Parameters. International Journal of Scientific Research in Science, Engineering and Technology. 2026;13(4):335-346. doi:10.32628/ijsrset26134131

BibTeX
@article{Kumbhar2026,
  author = {Kumbhar, N. N. and Tilekar, Dr. S. K.},
  title = {Design and Experimental Validation of a Mixed Signal PSoC Based System on Chip for Monitoring of Neonatal Incubator Parameters},
  journal = {International Journal of Scientific Research in Science, Engineering and Technology},
  year = {2026},
  volume = {13},
  number = {4},
  pages = {335--346},
  doi = {10.32628/ijsrset26134131}
}
Kostenlos starten →