EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

TracheoSense: Innovation in Endotracheal Tube Cuff Pressure Monitoring with a Three-Dimensional Printed Electronic Manometer-Syringe Device

Cloie Anne Rabinetas, Antonio Chua, Thomas Niccolo Reyes

Philippine Journal of Otolaryngology Head and Neck Surgery · 21.04.2026 · 41 (1) · 55

DOI: 10.32412/pjohns.v41i1.2383 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Philippine Journal of Otolaryngology Head and Neck Surgery, online erschienen am 21.04.2026. Verfasst von Cloie Anne Rabinetas und 2 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective:To design and validate TracheoSense, a 3D-printed manometer-syringe device for measuring endotracheal tube (ETT) cuff pressure, by assessing its accuracy and precision compared with a commercially available standard manometer. Methods: Design: Instrument Innovation Setting: Tertiary Private Training Hospital Participants: None Results:The study compared the TracheoSense device with the VBM Analog Manometer, the reference standard, for measuring cuff air volume at target pressures of 20 and 30 cmH2O across tracheal model diameters of 19 mm, 21 mm, and 23 mm. At 20 cmH₂O, mean cuff volumes measured by TracheoSense ranged from 4.04 ± 0.09 mL to 5.82 ± 0.07 mL, with coefficients of variation (CV) between 1.15% and 2.18%, and minimal bias compared to the reference device (−0.02 to 0.09 mL). At 30 cmH₂O, mean volumes ranged from 4.51 ± 0.11 mL to 6.31 ± 0.23 mL, with CVs of 1.62% to 3.58% and biases between −0.07 and 0.02 mL. Conclusion:The TracheoSense 3D-printed manometer-syringe device demonstrated comparable accuracy and precision in measuring ETT cuff pressures when compared with a commercially available analog manometer. These results support this device as a reliable and practical tool for cuff pressure monitoring.

Zitieren

APA

Rabinetas, C. A., Chua, A., & Reyes, T. N. (2026). TracheoSense: Innovation in Endotracheal Tube Cuff Pressure Monitoring with a Three-Dimensional Printed Electronic Manometer-Syringe Device. Philippine Journal of Otolaryngology Head and Neck Surgery, 41(1), 55. https://doi.org/10.32412/pjohns.v41i1.2383

Vancouver

Rabinetas CA, Chua A, Reyes TN. TracheoSense: Innovation in Endotracheal Tube Cuff Pressure Monitoring with a Three-Dimensional Printed Electronic Manometer-Syringe Device. Philippine Journal of Otolaryngology Head and Neck Surgery. 2026;41(1):55. doi:10.32412/pjohns.v41i1.2383

BibTeX
@article{Rabinetas2026,
  author = {Rabinetas, Cloie Anne and Chua, Antonio and Reyes, Thomas Niccolo},
  title = {TracheoSense: Innovation in Endotracheal Tube Cuff Pressure Monitoring with a Three-Dimensional Printed Electronic Manometer-Syringe Device},
  journal = {Philippine Journal of Otolaryngology Head and Neck Surgery},
  year = {2026},
  volume = {41},
  number = {1},
  pages = {55},
  doi = {10.32412/pjohns.v41i1.2383}
}