Abstract (Originalsprache)
The promontory test is applied to examine the functionality of the auditory nerve as part of the preoperative examination for cochlear implantation (CI). Besides standard tests, such as pure tone and speech audio-metry, imaging, vestibular tests and brainstem audiometry (BERA), the promontory test usually is applied in case of special issues (very long duration of deafness or other reasons for potential damage of the auditory nerve). The aim of the study is to evaluate the applicability of the freely programmable ISIS Neurostimulator (inomed Medizintechnik, Emmendingen) to perform the promontory test for preoperative examination for CI. Furthermore the possible effect of biphasic and triphasic pulses on the subjective perception is investigated. 29 patients (15 ♂, 14 ♀) with an age between 18 and 86 years were included in the study (35 ear/cases). An ear canal electrode (Medtronic) was positioned in saline solution in front of the eardrum (frontal reference electrode). For stimulation, biphasic and triphasic pulses with a repetition rate of 50 Hz were used. Via variation of the amplitude, the perception threshold and the uncomfortable level were determined for each pulse. In 10 out of 35 cases (28.6%) no hearing sensation but sensations of beginning pain was documented. The comparison of thresholds with biphasic and triphasic stimulation showed no significant difference regarding the perception threshold (median bi-/triphasic: 160/150 µA) and uncomfortable level (median bi-/triphasic: 320/315 µA). In contrast to findings obtained with triphasic pulses with intracochlear stimulation, the pulse shape had no impact on thresholds. In general, the results of the study show that the ISIS Neurostimulator is applicable as device for administering electrical stimulation with ear canal electrodes. The aim of further studies evaluating the ISIS Neurostimulator is the clinical testing of additional measuring parameters, such as variation in stimulus frequencies, as well as E-BERA tests.
Schlagwörter
Zitieren
APA
Bruschke, S., Wu, P.-P., Baumann, & U. (2019). ISIS Neurostimulator: Erste Erfahrungen bei bi- und triphasischer elektrischer Stimulation mittels Gehörgangselektrode. GMS Zeitschrift für Audiologie – Audiological Acoustics, 1-1. https://doi.org/10.3205/zaud000002
Vancouver
Bruschke, S., Wu, P.-P., Baumann, U.. ISIS Neurostimulator: Erste Erfahrungen bei bi- und triphasischer elektrischer Stimulation mittels Gehörgangselektrode. GMS Zeitschrift für Audiologie – Audiological Acoustics. 2019:1-1. doi:10.3205/zaud000002
BibTeX
@article{Bruschke2019,
author = {Bruschke and S. and Wu and P.-P. and Baumann and U.},
title = {ISIS Neurostimulator: Erste Erfahrungen bei bi- und triphasischer elektrischer Stimulation mittels Gehörgangselektrode},
journal = {GMS Zeitschrift für Audiologie – Audiological Acoustics},
year = {2019},
pages = {1--1},
doi = {10.3205/zaud000002}
}
EUVIMED