Impact of multipoint pelvic fixation and multirod distal constructs on proximal junction biomechanics in cadaveric specimens
Journal of Neurosurgery: Spine · 2025
DOI: 10.3171/2025.4.spine25263 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Neurosurgery: Spine, 2025. Verfasst von Robert F Rudy und 10 weiteren Autor:innen. Beteiligte Einrichtung: 1Department of Neurosurgery, Barrow Neurological Institute, St. Joseph's Hospital and Medical Center, Phoenix, Arizona. Hauptthemen laut MEDLINE: Spinal Fusion, Lumbar Vertebrae, Pelvis, Pelvic Bones. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 32 Literaturangaben.
Abstract (Originalsprache)
OBJECTIVE Multipoint pelvic fixation with multirod constructs is increasingly used for long-segment deformity constructs to reduce rates of distal failure. However, more robust distal fixation may negatively impact proximal junction biomechanics, and this potential relationship has not been extensively studied. METHODS Standard nondestructive flexibility tests (7.5 Nm) were performed on 7 cadaveric specimens (L1–pelvis) to assess intervertebral flexibility (range of motion [ROM]), rod strain, and screw bending moments along a posterior fusion construct (pedicle screw and rod [PSR]) spanning L2–S1, supplemented by bilateral primary S2 alar-iliac (S2AI) fixation (2 S2AI screws and 2 rods), followed by additional S2AI screw placement and bilateral accessory rod placement spanning L4–S2AI (4 S2AI screws and 4 rods). Four conditions were tested for each specimen: 1) intact; 2) L2–S1 PSR; 3) L2–S2AI PSR; and 4) L2–S2AI plus L4–S2AI. Data were analyzed using repeated-measures ANOVA. RESULTS Seven cadaveric specimens were included. Proximal rod strain at the L2–3 level did not change across the varying test conditions in the 7 specimens tested (p > 0.05 for all conditions). There was no significant difference detected in proximal screw strain across conditions (p > 0.05). Finally, no significant difference was found in L2–3 ROM (p > 0.05) across instrumented variations, all of which were more rigid than intact specimens. CONCLUSIONS Pelvic fixation with 2 or 4 screws and 2 or 4 rods, respectively, did not significantly alter proximal junction screw or rod strain in a cadaveric model. Robust pelvic fixation might protect against distal failure without deleterious effects on the proximal junction.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- 1Department of Neurosurgery, Barrow Neurological Institute, St. Joseph's Hospital and Medical Center, Phoenix, Arizona.
- 2SI-BONE, Santa Clara, California; and.
- 3Department of Orthopedic Surgery, University of Minnesota, Minneapolis, Minnesota.
Zitieren
APA
Rudy, R. F., Sawa, A. G. U., McBryan, S., Mugge, L. A., Thielen, K., Assefa, T. G., Lindsey, D. P., Polly, D. W., Uribe, J. S., Kelly, B. P., & Turner, J. D. (2025). Impact of multipoint pelvic fixation and multirod distal constructs on proximal junction biomechanics in cadaveric specimens. Journal of Neurosurgery: Spine. https://doi.org/10.3171/2025.4.spine25263
Vancouver
Rudy RF, Sawa AGU, McBryan S, Mugge LA, Thielen K, Assefa TG, et al. Impact of multipoint pelvic fixation and multirod distal constructs on proximal junction biomechanics in cadaveric specimens. Journal of Neurosurgery: Spine. 2025. doi:10.3171/2025.4.spine25263
BibTeX
@article{Rudy2025,
author = {Rudy, Robert F and Sawa, Anna G U and McBryan, Sarah and Mugge, Luke A and Thielen, Katherine and Assefa, Temesgen G and Lindsey, Derek P and Polly, David W and Uribe, Juan S and Kelly, Brian P and Turner, Jay D},
title = {Impact of multipoint pelvic fixation and multirod distal constructs on proximal junction biomechanics in cadaveric specimens},
journal = {Journal of Neurosurgery: Spine},
year = {2025},
doi = {10.3171/2025.4.spine25263}
}
EUVIMED