EUVIMEDEUVIMED entdecken
Crossref

Comparison of the Contour Neurovascular System and Woven EndoBridge device for treatment of wide-necked cerebral aneurysms at a bifurcation or sidewall

Constantin Hecker, Erasmia Broussalis, Johannes A R Pfaff, Slaven Pikija, Christoph J Griessenauer, Monika Killer-Oberpfalzer

Journal of Neurosurgery · 01.08.2023 · 139 (2) · 563-572

DOI: 10.3171/2022.12.jns222268 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Neurosurgery, erschienen am 01.08.2023. Verfasst von Constantin Hecker und 5 weiteren Autor:innen. Beteiligte Einrichtung: 1Institute of Neurointervention, Paracelsus Medical University, Salzburg; and. Hauptthemen laut MEDLINE: Intracranial Aneurysm, Embolization, Therapeutic, Endovascular Procedures, Thromboembolism. Die Arbeit wurde bisher 25-mal zitiert (Crossref) und nennt 33 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE The authors compared the Contour Neurovascular System (Contour) with the Woven EndoBridge (WEB) device for the treatment of wide-necked cerebral aneurysms at a bifurcation or sidewall. METHODS Prospective clinical and radiological data were collected for all patients treated with either the Contour or WEB at a tertiary university hospital from May 2018 to June 2022. RESULTS In patients who had at least 3 months of follow-up data available (median patient age 60.0 [IQR 51.8–67.0] years, male/female ratio 1:1.4), the authors compared 40 aneurysms in 34 patients treated with the Contour and 30 aneurysms in 30 patients treated with the WEB. Overall, 26 middle cerebral artery, 24 anterior communicating artery, 9 basilar artery tip, 4 posterior communicating artery, 4 internal carotid artery, 1 anterior cerebral artery, 1 posterior inferior cerebellar artery, and 1 superior cerebellar artery aneurysm were treated. In the Contour cohort, complete occlusion at last follow-up was achieved for 30 aneurysms (75%) and a small neck remnant was seen in 6 aneurysms (15%), summing up to an adequate occlusion rate of 90%. One aneurysm (2.5%) had to be retreated, and 1 symptomatic thromboembolic event (2.5%) was observed with complete remission at discharge. Three adjunctive stents (10%) had to be used due to branch occlusion. In the WEB cohort, adequate occlusion was also seen in 90% of aneurysms (complete occlusion in 19 [63.3%] and remnant neck in 8 [26.7%], with a retreatment rate of 20%). Four WEBs (13.3%) needed additional stent placement due to device protrusion into a branch, 2 asymptomatic thromboembolic events (6.7%) were noted, and 1 major ischemic event (3.3%) due to M2 occlusion was noted. One patient treated with the WEB died between follow-ups of causes unrelated to the aneurysm, treatment, or device. Time from first measurement to deployment and thus total treatment time was significantly shorter in the Contour group (p = 0.004), regardless of whether a prior angiogram was available for aneurysm measurement and device sizing. CONCLUSIONS Results for the Contour were promising, although longer follow-up is necessary to draw more solid conclusions on the utility and risk profile of this new device compared with the already widely used WEB device. Adequate occlusion at last follow-up was the same for both devices, whereas the probability of complete occlusion at last follow-up was significantly higher for the Contour, and the WEB showed a significantly higher retreatment rate. Median deployment times were significantly shorter with the Contour than the WEB.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Intracranial AneurysmEmbolization, TherapeuticEndovascular ProceduresThromboembolismHumansMaleFemaleMiddle AgedTreatment OutcomeProspective StudiesRetrospective Studies

Stichwörter der Autor:innen

WEBWoven EndoBridgeaneurysmcontourembolizationendovascular neurosurgerynew devicesurgical techniquevascular disorders

Autor:innen und Institutionen

  • 1Institute of Neurointervention, Paracelsus Medical University, Salzburg; and.
  • Departments of2Neurosurgery.
  • 3Neurology, and.
  • 4Neuroradiology, Paracelsus Medical University, Salzburg, Austria.

Zitieren

APA

Hecker, C., Broussalis, E., Pfaff, J. A. R., Pikija, S., Griessenauer, C. J., & Killer-Oberpfalzer, M. (2023). Comparison of the Contour Neurovascular System and Woven EndoBridge device for treatment of wide-necked cerebral aneurysms at a bifurcation or sidewall. Journal of Neurosurgery, 139(2), 563-572. https://doi.org/10.3171/2022.12.jns222268

Vancouver

Hecker C, Broussalis E, Pfaff JAR, Pikija S, Griessenauer CJ, Killer-Oberpfalzer M. Comparison of the Contour Neurovascular System and Woven EndoBridge device for treatment of wide-necked cerebral aneurysms at a bifurcation or sidewall. Journal of Neurosurgery. 2023;139(2):563-572. doi:10.3171/2022.12.jns222268

BibTeX
@article{Hecker2023,
  author = {Hecker, Constantin and Broussalis, Erasmia and Pfaff, Johannes A R and Pikija, Slaven and Griessenauer, Christoph J and Killer-Oberpfalzer, Monika},
  title = {Comparison of the Contour Neurovascular System and Woven EndoBridge device for treatment of wide-necked cerebral aneurysms at a bifurcation or sidewall},
  journal = {Journal of Neurosurgery},
  year = {2023},
  volume = {139},
  number = {2},
  pages = {563--572},
  doi = {10.3171/2022.12.jns222268}
}
Kostenlos starten →