Survey and Analysis of Current State of Ventilator Alarms in the Intensive Care Unit
Global Clinical Engineering Journal · 2018
DOI: 10.31354/globalce.v0i1.25 ↗Kurz gefasst
Fachartikel in der Zeitschrift Global Clinical Engineering Journal, 2018. Verfasst von Z K Lin und 1 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Themen: Intensivpflege
Abstract (Originalsprache)
This article reports on a survey and analysis of ventilator alarm state in a children hospital. Based on the evaluation of the alarm effectiveness, we designed a survey statistical table for ventilator alarm investigation. We evaluated the alarm situation synthetically through investigation and statistical methods. Result shows that the current ventilator alarms are not sufficiently effective, 26.84% of them are meaningless alarms and those leading to clinician’s intervention make up only 2.26% of all the alarms generated. The reliability of statistical data was also analyzed. According to the survey results, we identified and analyzed the causes of the problem and proposed the corresponding alarm management methods.
Zitieren
APA
Lin, Z. K., & Zheng, K. (2018). Survey and Analysis of Current State of Ventilator Alarms in the Intensive Care Unit. Global Clinical Engineering Journal. https://doi.org/10.31354/globalce.v0i1.25
Vancouver
Lin ZK, Zheng K. Survey and Analysis of Current State of Ventilator Alarms in the Intensive Care Unit. Global Clinical Engineering Journal. 2018. doi:10.31354/globalce.v0i1.25
BibTeX
@article{Lin2018,
author = {Lin, Z K and Zheng, Kun},
title = {Survey and Analysis of Current State of Ventilator Alarms in the Intensive Care Unit},
journal = {Global Clinical Engineering Journal},
year = {2018},
doi = {10.31354/globalce.v0i1.25}
}