EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Carotid Artery Corrected Flow Time Measured by Wearable Doppler Ultrasound Accurately Detects Changing Stroke Volume During the Passive Leg Raise in Ambulatory Volunteers

Jon-Émile S Kenny, Christine Horner, Mai Elfarnawany, Andrew M Eibl, Joseph K Eibl

Frontiers in Bioscience-Elite · 22.05.2023 · 15 (2)

DOI: 10.31083/j.fbe1502012 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Bioscience-Elite, online erschienen am 22.05.2023. Verfasst von Jon-Émile S Kenny und 4 weiteren Autor:innen. Beteiligte Einrichtung: Health Sciences North Research Institute, Sudbury, ON P3E 2H2, Canada. Hauptthemen laut MEDLINE: Leg, Wearable Electronic Devices. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The change in the corrected flow time of the common carotid artery (ccFTΔ) has been used as a surrogate of changing stroke volume (SVΔ) in the critically-ill. Thus, this relatively easy-to-obtain Doppler measure may help clinicians better define the intended effect of intravenous fluids. Yet the temporal evolution of SVΔ and ccFTΔ has not been reported in volunteers undergoing a passive leg raise (PLR).

Methods

We recruited clinically-euvolemic, non-fasted, adult, volunteers in a local physiology lab to perform 2 PLR maneuvers, each separated by a 5 minute ‘wash-out’. During each PLR, SV was measured by a non-invasive pulse contour analysis device. SV was temporally-synchronized with a wireless, wearable Doppler ultrasound worn over the common carotid artery that continuously measured ccFT.

Results

36 PLR maneuvers were obtained across 19 ambulatory volunteers. 8856 carotid Doppler cardiac cycles were analyzed. The ccFT increased nearly ubiquitously during the PLR and within 40–60 seconds of PLR onset; the rise in SV from the pulse contour device was more gradual. SVΔ by +5% and +10% were both detected by a +7% ccFTΔ with sensitivities, specificities and areas under the receiver operator curve of 59%, 95% and 0.77 (p < 0.001) and 66%, 76% and 0.73 (p < 0.001), respectively.

Conclusions

The ccFTΔ during the PLR in ambulatory volunteers was rapid and sustained. Within the limits of precision for detecting a clinically-significant rise in SV by a non-invasive pulse contour analysis device, simultaneously-acquired ccFT from a wireless, wearable ultrasound system was accurate at detecting ‘preload responsiveness’.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

LegWearable Electronic DevicesAdultHumansStroke VolumeProspective StudiesCarotid ArteriesCarotid Artery, CommonUltrasonography, DopplerVolunteersHemodynamics

Stichwörter der Autor:innen

Frank-Starlingcarotid arterycorrected flow timefluid responsivenesspassive leg raise

Autor:innen und Institutionen

  • Health Sciences North Research Institute, Sudbury, ON P3E 2H2, Canada.
  • Flosonics Medical, Toronto, ON M5V 2Y1, Canada.
  • Northern Ontario School of Medicine, Sudbury, ON P3E 2C6, Canada.

Zitieren

APA

Kenny, J. S., Horner, C., Elfarnawany, M., Eibl, A. M., & Eibl, J. K. (2023). Carotid Artery Corrected Flow Time Measured by Wearable Doppler Ultrasound Accurately Detects Changing Stroke Volume During the Passive Leg Raise in Ambulatory Volunteers. Frontiers in Bioscience-Elite, 15(2). https://doi.org/10.31083/j.fbe1502012

Vancouver

Kenny JS, Horner C, Elfarnawany M, Eibl AM, Eibl JK. Carotid Artery Corrected Flow Time Measured by Wearable Doppler Ultrasound Accurately Detects Changing Stroke Volume During the Passive Leg Raise in Ambulatory Volunteers. Frontiers in Bioscience-Elite. 2023;15(2). doi:10.31083/j.fbe1502012

BibTeX
@article{Kenny2023,
  author = {Kenny, Jon-Émile S and Horner, Christine and Elfarnawany, Mai and Eibl, Andrew M and Eibl, Joseph K},
  title = {Carotid Artery Corrected Flow Time Measured by Wearable Doppler Ultrasound Accurately Detects Changing Stroke Volume During the Passive Leg Raise in Ambulatory Volunteers},
  journal = {Frontiers in Bioscience-Elite},
  year = {2023},
  volume = {15},
  number = {2},
  doi = {10.31083/j.fbe1502012}
}
Kostenlos starten →