Method Verification for The Determination of Cu and Zn Metal Ions in Seawater using Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES)
JURNAL KIMIA MULAWARMAN · 31.05.2026 · 23 (2) · 60-69
DOI: 10.30872/jkm.v23i2.1487 ↗Kurz gefasst
Fachartikel in der Zeitschrift JURNAL KIMIA MULAWARMAN, online erschienen am 31.05.2026. Verfasst von Adinda Anno Meitasari und 2 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY-SA 4.0).
Abstract (Originalsprache)
Research on verification of the method for determining Cu and Zn metal ions using ICP-OES for seawater samples, namely AL-1 and AL-2, has been carried out. The verification value for Cu metal is Linearity = 1,000; Accuracy (% trueness) = 100.7; % bias = -0.74; Precision (% RSD = 1.47 ≤ 2/3 % CV Horwitz = 16.82); % Recovery = 100.2%; MDL = 0.01; LOQ = 0.05; Uncertainty value = 0.03; Low repeatability (% RSD = 2.17 < 0.5 CV Horwitz = 7.13); middle (% RSD = 0.67 < 0.5 CV Horwitz = 6.72); high (% RSD = 0.72 < 0.5 CV Horwitz = 5.85); Reproducibility (horrat value) low = 0.58; middle = 0.38 and high = 0.34. The verification value of Zn metal is Linearity = 1,000; Accuracy (% trueness) = 95.1; % bias = 4.93; Precision (% RSD = 10.10 ≤ 2/3 % CV Horwitz = 16.94); % Recovery = 95.6%; MDL = 0.01; LOQ = 0.05; Uncertainty value = 0.04. Low repeatability (% RSD = 1.17 < 0.5 CV Horwitz = 7.17); middle (% RSD = 0.69 < 0.5 CV Horwitz = 6.28); high (% RSD = 0.70 < 0.5 CV Horwitz = 5.85). Reproducibility (horrat value) low concentration= 0.57; medium concentration = 0.50; high concentration = 0.33. The sample concentrations and uncertainties of Cu and Zn metals are AL-1 (0.25 ± 0.03) and (0.01 ± 0.04); AL-2 (0.30 ± 0.03) and (0.40 ± 0.04). The measurement results show that the verification parameter values are requirements for determining metals in seawater.
Zitieren
APA
Meitasari, A. A., Panggabean, P. D. A. S., & M.Si, P. D. S. P. P. (2026). Method Verification for The Determination of Cu and Zn Metal Ions in Seawater using Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES). JURNAL KIMIA MULAWARMAN, 23(2), 60-69. https://doi.org/10.30872/jkm.v23i2.1487
Vancouver
Meitasari AA, Panggabean PDAS, M.Si PDSPP. Method Verification for The Determination of Cu and Zn Metal Ions in Seawater using Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES). JURNAL KIMIA MULAWARMAN. 2026;23(2):60-69. doi:10.30872/jkm.v23i2.1487
BibTeX
@article{Meitasari2026,
author = {Meitasari, Adinda Anno and Panggabean, Prof. Dr. Aman S and M.Si, Prof. Dr. Subur P Pasaribu},
title = {Method Verification for The Determination of Cu and Zn Metal Ions in Seawater using Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES)},
journal = {JURNAL KIMIA MULAWARMAN},
year = {2026},
volume = {23},
number = {2},
pages = {60--69},
doi = {10.30872/jkm.v23i2.1487}
}