Characterization of an evanescent wave-fiber optic sensor for formalin detection

Widayanti, Anindita R A

Compton: Jurnal Ilmiah Pendidikan Fisika · 11.07.2026 · 12 (2) · 217-229

DOI: 10.30738/cjipf.v12i2.22164 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Compton: Jurnal Ilmiah Pendidikan Fisika, online erschienen am 11.07.2026. Verfasst von Widayanti und 1 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

It is necessary to design a fast, inexpensive, and portable method for detecting formalin to support food safety screening and halal-tayyib principles. This study aims to develop and characterize a formalin detection system using a smartphone, an optical fiber sensor, an evanescent wave, and a diffraction grating. The sensor medium comprises a multi-mode step-index plastic-clad silica optical fiber (core diameter 400 µm; NA 0.48), with modifications in the sensing area to enhance interaction between the evanescent field and the medium. The optical fiber used in the detection system has its end polished to a 45-degree angle and its cladding section scratched at five locations over a 1 cm length. The optical system uses the smartphone’s rear LED as a light source, the camera as a detector, and a diffraction grating from a DVD to disperse the output light, enabling intensity changes to be recorded. The spectral image is analyzed in ImageJ to obtain the intensities of the R, G, and B channels, then converted to a single grayscale intensity value using a weighted luminance model and evaluated using linear regression. The results show that RGB and grayscale intensities increase monotonically with formalin concentration from 0 to 10 ppm. The grayscale calibration curve is given by the equation. . The sensor's sensitivity indicates that a 1 ppm change in concentration corresponds to a 1.7481 pixel change. The regression value (R2) of 0.9565 demonstrates good linearity across the test range. Per-channel analysis indicates that the green channel exhibits the highest linearity, whereas the blue channel shows the greatest response. These findings confirm that the proposed optical fiber-smartphone sensor platform has the potential to be a compact, low-cost, and viable solution for portable formalin detection. It is necessary to design a fast, inexpensive, and portable method for detecting formalin to support food safety screening and halal-tayyib principles. This study aims to develop and characterize a formalin detection system using a smartphone, an optical fiber sensor, an evanescent wave, and a diffraction grating. The sensor medium comprises a multi-mode step-index plastic-clad silica optical fiber (core diameter 400 µm; NA 0.48), with modifications in the sensing area to enhance interaction between the evanescent field and the medium. The optical fiber used in the detection system has its end polished to a 45-degree angle and its cladding section scratched at five locations over a 1 cm length. The optical system uses the smartphone’s rear LED as a light source, the camera as a detector, and a diffraction grating from a DVD to disperse the output light, enabling intensity changes to be recorded. The spectral image is analyzed in ImageJ to obtain the intensities of the R, G, and B channels, then converted to a single grayscale intensity value using a weighted luminance model and evaluated using linear regression. The results show that RGB and grayscale intensities increase monotonically with formalin concentration from 0 to 10 ppm. The grayscale calibration curve is given by the equation. . The sensor's sensitivity indicates that a 1 ppm change in concentration corresponds to a 1.7481 pixel change. The regression value (R2) of 0.9565 demonstrates good linearity across the test range. Per-channel analysis indicates that the green channel exhibits the highest linearity, whereas the blue channel shows the greatest response. These findings confirm that the proposed optical fiber-smartphone sensor platform has the potential to be a compact, low-cost, and viable solution for portable formalin detection.

Zitieren

APA

Widayanti, & A, A. R. (2026). Characterization of an evanescent wave-fiber optic sensor for formalin detection. Compton: Jurnal Ilmiah Pendidikan Fisika, 12(2), 217-229. https://doi.org/10.30738/cjipf.v12i2.22164

Vancouver

Widayanti, A AR. Characterization of an evanescent wave-fiber optic sensor for formalin detection. Compton: Jurnal Ilmiah Pendidikan Fisika. 2026;12(2):217-229. doi:10.30738/cjipf.v12i2.22164

BibTeX
@article{Widayanti2026,
  author = {Widayanti and A, Anindita R},
  title = {Characterization of an evanescent wave-fiber optic sensor for formalin detection},
  journal = {Compton: Jurnal Ilmiah Pendidikan Fisika},
  year = {2026},
  volume = {12},
  number = {2},
  pages = {217--229},
  doi = {10.30738/cjipf.v12i2.22164}
}
Kostenlos starten →