An International Study of Factors Affecting Variability of Dosimetry Calculations, Part 3: Contribution from Calculating Absorbed Dose from Time-Integrated Activity
Journal of Nuclear Medicine · 2024-08 · 65 (8) · 1166-1172
DOI: 10.2967/jnumed.123.267293 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Nuclear Medicine, online erschienen am 03.07.2024. Verfasst von Julia Brosch-Lenz und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nuclear Medicine, Rechts der Isar Medical Center, Technical University of Munich, Munich, Germany. Hauptthemen laut MEDLINE: Radiometry. Die Arbeit wurde bisher 27-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NCI NIH HHS.
Abstract (Originalsprache)
Image-based dosimetry-guided radiopharmaceutical therapy has the potential to personalize treatment by limiting toxicity to organs at risk and maximizing the therapeutic effect. The 177Lu dosimetry challenge of the Society of Nuclear Medicine and Molecular Imaging consisted of 5 tasks assessing the variability in the dosimetry workflow. The fifth task investigated the variability associated with the last step, dose conversion, of the dosimetry workflow on which this study is based.
Methods
Reference variability was assessed by 2 medical physicists using different software, methods, and all possible combinations of input segmentation formats and time points as provided in the challenge. General descriptive statistics for absorbed dose values from the global submissions from participants were calculated, and variability was measured using the quartile coefficient of dispersion.
Results
For the liver, which included lesions with high uptake, variabilities of up to 36% were found. The baseline analysis showed a variability of 29% in absorbed dose results for the liver from datasets where lesions included and excluded were grouped, indicating that variation in how lesions in normal liver were treated was a significant source of variability. For other organs and lesions, variability was within 7%, independently of software used except for the local deposition method.
Conclusion
The choice of dosimetry method or software had a small contribution to the overall variability of dose estimates.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Nuclear Medicine, Rechts der Isar Medical Center, Technical University of Munich, Munich, Germany.
- Department of Nuclear Medicine, Pomeranian Medical University, Szczecin, Poland.
- Department of Integrative Oncology, BC Cancer Research Institute, Vancouver, British Columbia, Canada.
- Rapid, LLC, Baltimore, Maryland.
- Department of Radiology, Johns Hopkins University, Baltimore, Maryland.
- Department of Radiology, University of Michigan, Ann Arbor, Michigan.
Förderung
- NCI NIH HHS
Zitieren
APA
Brosch-Lenz, J., Kurkowska, S., Frey, E., Dewaraja, Y. K., Sunderland, J., & Uribe, C. (2024). An International Study of Factors Affecting Variability of Dosimetry Calculations, Part 3: Contribution from Calculating Absorbed Dose from Time-Integrated Activity. Journal of Nuclear Medicine, 65(8), 1166-1172. https://doi.org/10.2967/jnumed.123.267293
Vancouver
Brosch-Lenz J, Kurkowska S, Frey E, Dewaraja YK, Sunderland J, Uribe C. An International Study of Factors Affecting Variability of Dosimetry Calculations, Part 3: Contribution from Calculating Absorbed Dose from Time-Integrated Activity. Journal of Nuclear Medicine. 2024;65(8):1166-1172. doi:10.2967/jnumed.123.267293
BibTeX
@article{BroschLenz2024,
author = {Brosch-Lenz, Julia and Kurkowska, Sara and Frey, Eric and Dewaraja, Yuni K and Sunderland, John and Uribe, Carlos},
title = {An International Study of Factors Affecting Variability of Dosimetry Calculations, Part 3: Contribution from Calculating Absorbed Dose from Time-Integrated Activity},
journal = {Journal of Nuclear Medicine},
year = {2024},
volume = {65},
number = {8},
pages = {1166--1172},
doi = {10.2967/jnumed.123.267293}
}