Image Quality of Cardiac Silicon Photomultiplier PET/CT Using an Infant Phantom of Extremely Low Birth Weight
Journal of Nuclear Medicine Technology · 2024
DOI: 10.2967/jnmt.124.267826 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Nuclear Medicine Technology, 2024. Verfasst von Kazuki Fukuchi und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Medical Physics and Engineering, Course of Health Science, Osaka University Graduate School of Medicine, Osaka, Japan; kfukuchi@sahs.me…. Hauptthemen laut MEDLINE: Phantoms, Imaging, Positron Emission Tomography Computed Tomography, Heart, Silicon. Die Arbeit nennt 23 Literaturangaben.
Abstract (Originalsprache)
The lack of pediatrics-specific equipment for nuclear medicine imaging has resulted in insufficient diagnostic information for newborns, especially low-birth-weight infants. Although PET offers high spatial resolution and low radiation exposure, its use in newborns is limited. This study investigated the feasibility of cardiac PET imaging using the latest silicon photomultiplier (SiPM) PET technology in infants of extremely low birth weight (ELBW) using a phantom model.
Methods
The study used a phantom model representing a 500-g ELBW infant with brain, cardiac, liver, and lung tissues. The cardiac tissue included a 3-mm-thick defect mimicking myocardial infarction. Organ tracer concentrations were calculated assuming 18F-FDG myocardial viability scans and 18F-flurpiridaz myocardial perfusion scans and were added to the phantom organs. Imaging was performed using an SiPM PET/CT scanner with a 5-min acquisition. The data acquired in list mode were reconstructed using 3-dimensional ordered-subsets expectation maximization with varying iterations. Image evaluation was based on the depiction of the myocardial defect compared with normal myocardial accumulation.
Results
Increasing the number of iterations improved the contrast of the myocardial defect for both tracers, with 18F-flurpiridaz showing higher contrast than 18F-FDG. However, even at 50 iterations, both tracers overestimated the defect accumulation. A bull's-eye image can display the flow metabolism mismatch using images from both tracers.
Conclusion
SiPM PET enabled cardiac PET imaging in a 500-g ELBW phantom with a 1-g heart. However, there were limitations in adequately depicting these defects. Considering the image quality and defect contrast,18F-flurpiridaz appears more desirable than 18F-FDG if only one of the two can be used.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Medical Physics and Engineering, Course of Health Science, Osaka University Graduate School of Medicine, Osaka, Japan; kfukuchi@sahs.med.osaka-u.ac.jp.
- Department of Quantum Medical Technology, Institute of Medical, Pharmaceutical, and Health Sciences, Kanazawa University, Kanazawa, Japan; and.
- Department of Radiology, National Cerebral and Cardiovascular Center, Osaka, Japan.
Zitieren
APA
Fukuchi, K., Shibutani, T., Terakawa, Y., Nouno, Y., Tateishi, E., Onoguchi, M., & Tetsuya, F. (2024). Image Quality of Cardiac Silicon Photomultiplier PET/CT Using an Infant Phantom of Extremely Low Birth Weight. Journal of Nuclear Medicine Technology. https://doi.org/10.2967/jnmt.124.267826
Vancouver
Fukuchi K, Shibutani T, Terakawa Y, Nouno Y, Tateishi E, Onoguchi M, et al. Image Quality of Cardiac Silicon Photomultiplier PET/CT Using an Infant Phantom of Extremely Low Birth Weight. Journal of Nuclear Medicine Technology. 2024. doi:10.2967/jnmt.124.267826
BibTeX
@article{Fukuchi2024,
author = {Fukuchi, Kazuki and Shibutani, Takayuki and Terakawa, Yusuke and Nouno, Yoshifumi and Tateishi, Emi and Onoguchi, Masahisa and Tetsuya, Fukuda},
title = {Image Quality of Cardiac Silicon Photomultiplier PET/CT Using an Infant Phantom of Extremely Low Birth Weight},
journal = {Journal of Nuclear Medicine Technology},
year = {2024},
doi = {10.2967/jnmt.124.267826}
}
EUVIMED