Quantification of the pressures generated during insertion of an epidural needle in labouring women of varying body mass indices
International Journal of Clinical Anesthesia and Research · 2017
DOI: 10.29328/journal.hacr.1001004 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Clinical Anesthesia and Research, 2017. Verfasst von Wee MYK und 4 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective
The primary aim of this study was to measure pressure generated on a Tuohy needle during the epidural procedure in labouring women of varying body mass indices (BMI) with a view of utilising the data for the future development of a high fidelity epidural simulator. High-fidelity epidural simulators have a role in improving training and safety but current simulators lack a realistic experience and can be improved.
Methods
This study was approved by the National Research Ethics Service Committee South Central, Portsmouth (REC reference 11/SC/0196). After informed consent epidural needle insertion pressure was measured using a Portex 16-gauge Tuohy needle, loss-of-resistance syringe, a three-way tap, pressure transducer and a custom-designed wireless transmitter. This was performed in four groups of labouring women, stratified according to BMI kg/m2: 18-24.9; 25-34.9; 35-44.9 and >=45. One-way ANOVA was used to compare difference in needle insertion pressure between the BMI groups. A paired t-test was performed between BMI group 18-24.9 and the three other BMI groups. Ultrasound images of the lumbar spine were undertaken prior to the epidural procedure and lumbar magnetic resonance imaging (MRI) was performed within 72h post-delivery. These images will be used in the development of a high fidelity epidural simulator.
Results
The mean epidural needle insertion pressure of labouring women with BMI 18-24.9 was 461mmHg; BMI 25-34.9 was 430mmHg; BMI 35-44.9 was 415mmHg and BMI >=45 was 376mmHg, (p=0.52).
Conclusion
Although statistically insignificant, the study did show a decreasing trend of epidural insertion pressure with increasing body mass indices.
Zitieren
APA
MYK, W., RA, I., N, V., VN, D., & B, P. (2017). Quantification of the pressures generated during insertion of an epidural needle in labouring women of varying body mass indices. International Journal of Clinical Anesthesia and Research. https://doi.org/10.29328/journal.hacr.1001004
Vancouver
MYK W, RA I, N V, VN D, B P. Quantification of the pressures generated during insertion of an epidural needle in labouring women of varying body mass indices. International Journal of Clinical Anesthesia and Research. 2017. doi:10.29328/journal.hacr.1001004
BibTeX
@article{MYK2017,
author = {MYK, Wee and RA, Isaacs and N, Vaughan and VN, Dubey and B, Parker},
title = {Quantification of the pressures generated during insertion of an epidural needle in labouring women of varying body mass indices},
journal = {International Journal of Clinical Anesthesia and Research},
year = {2017},
doi = {10.29328/journal.hacr.1001004}
}
EUVIMED