EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Influence of the biomechanical behavior of TiO2 and ZrO2 implants after implantoplasty.

Jefferson D. M. de Matos, Mário A. C. Sinhoreti, Nathália de C. Ramos, Daher A. Queiroz, Adriano B. Barbosa, Edinardo F. F. Matias, Thiago R. V. Viaro, Eric Padovani, Jerry D. Oliveira, Guilherme M. Rabelo, Marco A. Bottino, Guilherme da R. S. Lopes

Kariri Science - CECAPE Biology and Health Journal · 2023 · 1 (2) · 1-9

DOI: 10.29327/2256856.1.2-10 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Kariri Science - CECAPE Biology and Health Journal, 2023. Verfasst von Jefferson D. M. de Matos und 11 weiteren Autor:innen. Die Arbeit nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Peri-implantitis occurs around dental implants, and implantoplasty has been used to address this ongoing disease; however, the changes to the physical properties of an implant after implantoplasty have not been well documented. This in vitro study aimed to investigate the effect of implantoplasty on fracture strength and the load required for plastic deformation after fatigue cyclic. Twenty Yt-tria-stabilized Zirconia implants (ZrO2 group; Ø 4.1 x 10 mm, Straumann) and commercially pure titanium implants (TiO2 group; Ø 4.1 x 10 mm, Straumann, control group) implants were placed in polyurethane blocks with 5mm supracrestal. After that, 10 Zirconia and 10 Titanium implants were submitted to implantoplasty in the supracrestal portion using a high-speed handpiece with dia-mond burs. Then 4 groups (n=10) based on material and implantoplasty treatment were: Ti-Before (titanium implant control group), Zr-Before (zirconia implant control group), Ti-After (titanium implant test group with implantoplasty), and Zr-After (zirconia implant test group with implan-toplasty). Each implant received its respective titanium abutment system (Purebase or Variobase), installed with a torque of 35 N.cm using a digital torque wrench. The specimens of each group were submitted to step stress test (30°; 10 Hz; 2 million cycles) until failure and the fracture strength (FS) values were recorded. Two-way ANOVA and parametric comparisons with control were adopted. The FS values for Ti-Bef, Zr-Bef, Ti-Aft, and Zr-Aft were 703.6 ± 68.1 N, 1225 ± 123.8 N, 956.6 ± 86.4 N, and 1035.1 ± 85.7 N, respectively. The number of cycles for Ti-Bef, Zr-Bef, Ti-Aft, and Zr-Aft was 379,323 ± 9,354; 564,779 ± 29,903; 351,451 ± 9,904; 467,009 ± 17,641, respectively. The Ti-Bef and Zr-Bef groups showed a significant FS and a number of cycles than the test groups (P

Zitieren

APA

Matos, J. D. M. D., Sinhoreti, M. A. C., Ramos, N. D. C., Queiroz, D. A., Barbosa, A. B., Matias, E. F. F., Viaro, T. R. V., Padovani, E., Oliveira, J. D., Rabelo, G. M., Bottino, M. A., & Lopes, G. D. R. S. (2023). Influence of the biomechanical behavior of TiO2 and ZrO2 implants after implantoplasty. Kariri Science - CECAPE Biology and Health Journal, 1(2), 1-9. https://doi.org/10.29327/2256856.1.2-10

Vancouver

Matos JDMD, Sinhoreti MAC, Ramos NDC, Queiroz DA, Barbosa AB, Matias EFF, et al. Influence of the biomechanical behavior of TiO2 and ZrO2 implants after implantoplasty. Kariri Science - CECAPE Biology and Health Journal. 2023;1(2):1-9. doi:10.29327/2256856.1.2-10

BibTeX
@article{Matos2023,
  author = {Matos, Jefferson D. M. de and Sinhoreti, Mário A. C. and Ramos, Nathália de C. and Queiroz, Daher A. and Barbosa, Adriano B. and Matias, Edinardo F. F. and Viaro, Thiago R. V. and Padovani, Eric and Oliveira, Jerry D. and Rabelo, Guilherme M. and Bottino, Marco A. and Lopes, Guilherme da R. S.},
  title = {Influence of the biomechanical behavior of TiO2 and ZrO2 implants after implantoplasty},
  journal = {Kariri Science - CECAPE Biology and Health Journal},
  year = {2023},
  volume = {1},
  number = {2},
  pages = {1--9},
  doi = {10.29327/2256856.1.2-10}
}