Quantifying discrepancy between indicated and actual oxygen flow rates delivered by Comweld Ezi-flow low and standard flowmeters under hyperbaric conditions: a technical report

Yoav Aufgang, Bridget Devaney iD, Jason Chan, Ian Millar, Theo Tsouras

Diving and Hyperbaric Medicine Journal · 30.06.2026 · 56 (2) · 154-160

DOI: 10.28920/dhm56.2.154-160 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diving and Hyperbaric Medicine Journal, online erschienen am 30.06.2026. Verfasst von Yoav Aufgang und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Intensive Care and Hyperbaric Medicine, Alfred Health, Melbourne, Australia. Hauptthemen laut MEDLINE: Oxygen, Hyperbaric Oxygenation, Flowmeters. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

During clinical use of extracorporeal membrane oxygenation (ECMO) in hyperbaric conditions at our centre, upward titration of indicated sweep gas flow rates is required to maintain adequate CO2 clearance. This project measured the impact of hyperbaric pressure on oxygen flow rates delivered by the Comweld Ezi-Flow flowmeters used in our centre.

Methods

Oxygen flow rates through Comweld Ezi-flow standard and low oxygen gas flowmeters were set at 101.3 kPa (1 atmosphere absolute [atm abs]) and then measured at intervals up to 284 kPa (2.8 atm abs) using a calibrated gas flow analyser, with cross verification against a small Douglas bag test type apparatus. During testing, the chamber was compressed and decompressed at a rate of 10 kPa·min-1. Flow rates during chamber compression and decompression were compared.

Results

The indicated rate of oxygen gas flow through the unadjusted flowmeters changed minimally – typically rising by a maximum of half of the diameter of the indicator ball. The actual (volumetric) flow, tested across indicated flow rates from 3 to 12 L·min-1, was consistently reduced by approximately 50% as the chamber pressure increased from 101.3 to 284 kPa (1 to 2.8 atm abs). A slightly smaller reduction was observed when assessing the low flowmeter across the same pressure range; reductions of 40.0 and 43.3% were demonstrated at 0.3 to 0.6 L·min-1 respectively. Differences in flow rates between compression and decompression were minor except at the very lowest flows.

Conclusions

At 284 kPa (2.8 atm abs), actual volumetric flow of oxygen through Comweld Ezi-Flow flowmeters is dramatically reduced and this needs appropriate compensation to ensure therapeutic aims are achieved.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

OxygenHyperbaric OxygenationFlowmetersDiving and Hyperbaric MedicineHumansAtmospheric PressureCalibrationDecompressionCarbon DioxideEquipment Design

Stichwörter der Autor:innen

Flow dynamicsGas flowHyperbaric environmentHyperbaric oxygen treatmentMedical equipment

Substanzen

OxygenCarbon Dioxide

Autor:innen und Institutionen

  • Department of Intensive Care and Hyperbaric Medicine, Alfred Health, Melbourne, Australia.
  • School of Translational Medicine, Faculty of Medicine Nursing and Health Sciences, Monash University, Melbourne, Australia.
  • Emergency and Trauma Centre, Alfred Health, Melbourne, Australia.
  • School of Public Health and Preventive Medicine, Faculty of Medicine Nursing and Health Sciences, Monash University, Melbourne, Australia.
  • Corresponding author: Dr Bridget Devaney, Department of Intensive Care and Hyperbaric Medicine, Alfred Health, 55 Commercial Road, Melbourne, VIC 3004, Australia, b.devaney@alfred.org.au.

ORCID-Profile: Bridget Devaney

Zitieren

APA

Aufgang, Y., Devaney, B., Chan, J., Millar, I., & Tsouras, T. (2026). Quantifying discrepancy between indicated and actual oxygen flow rates delivered by Comweld Ezi-flow low and standard flowmeters under hyperbaric conditions: a technical report. Diving and Hyperbaric Medicine Journal, 56(2), 154-160. https://doi.org/10.28920/dhm56.2.154-160

Vancouver

Aufgang Y, Devaney B, Chan J, Millar I, Tsouras T. Quantifying discrepancy between indicated and actual oxygen flow rates delivered by Comweld Ezi-flow low and standard flowmeters under hyperbaric conditions: a technical report. Diving and Hyperbaric Medicine Journal. 2026;56(2):154-160. doi:10.28920/dhm56.2.154-160

BibTeX
@article{Aufgang2026,
  author = {Aufgang, Yoav and Devaney, Bridget and Chan, Jason and Millar, Ian and Tsouras, Theo},
  title = {Quantifying discrepancy between indicated and actual oxygen flow rates delivered by Comweld Ezi-flow low and standard flowmeters under hyperbaric conditions: a technical report},
  journal = {Diving and Hyperbaric Medicine Journal},
  year = {2026},
  volume = {56},
  number = {2},
  pages = {154--160},
  doi = {10.28920/dhm56.2.154-160}
}
Kostenlos starten →