Reliability and Agreement of Hand-Held Dynamometry Using Three Standard Rater Test Positions
International Journal of Sports Physical Therapy · 01.02.2025 · 20 (2)
DOI: 10.26603/001c.128286 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Sports Physical Therapy, online erschienen am 01.02.2025. Verfasst von Frank Aerts und 6 weiteren Autor:innen. Beteiligte Einrichtung: Doctor of Physical Therapy Program Manchester University. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background
The use of portable hand-held dynamometers is increasing in popularity due to their ease of use in different clinical settings, convenient size, portability, and overall affordability. Reported reliability for external fixation and rater-stabilized hand-held dynamometry (HHD) strength measurements have been found to be ‘good’ to ‘excellent’. Inconsistent agreement has been found between the two stabilization methods and isokinetic HHD testing. Purpose / Hypothesis Determine the reliability and agreement of HHD measurements in three different rater test positions against three different mechanically produced force magnitudes. The study compared measurements obtained by rater-stabilization to external fixation methods. Study Design Intra-rater and inter-rater reliability study.
Methods
Ten raters took measurements in three different rater test positions against three different force magnitudes created by an external force. Raters were blinded to the randomized force magnitudes. The rater’s measurements were compared to measurements taken against an external fixation stabilization device. To establish reliability, Intraclass Correlation Coefficient (ICC), and Standard Error of Measurement (SEM), and Minimal Detectable Change (MDC) were used. To establish agreement, error rates between the rater-stabilized and external fixation stabilization measurements were calculated.
Results
ICC’s were found to be ‘excellent’ at .97 and above. The relative SEM ranged from 0.2% to 0.9 % and the relative MDC ranged from 0.7% to 2.8%. The overall error rate was 15.5% and was influenced by force magnitude.
Conclusion
The use of standardized rater test positions resulted in ‘excellent’ intra-rater, inter-rater reliability, low SEM, and low MDC for rater-stabilized HHD measurements. A systematic error was observed, with rater-stabilized measurements resulting in higher values compared with values obtained with the external fixation method. Level of Evidence 3
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Doctor of Physical Therapy Program Manchester University.
Zitieren
APA
Aerts, F., Sheets, H., Anderson, C., Bussie, N., Hoskins, R., Maninga, A., & Novak, E. (2025). Reliability and Agreement of Hand-Held Dynamometry Using Three Standard Rater Test Positions. International Journal of Sports Physical Therapy, 20(2). https://doi.org/10.26603/001c.128286
Vancouver
Aerts F, Sheets H, Anderson C, Bussie N, Hoskins R, Maninga A, et al. Reliability and Agreement of Hand-Held Dynamometry Using Three Standard Rater Test Positions. International Journal of Sports Physical Therapy. 2025;20(2). doi:10.26603/001c.128286
BibTeX
@article{Aerts2025,
author = {Aerts, Frank and Sheets, Holly and Anderson, Chance and Bussie, Natalie and Hoskins, Rose and Maninga, Amanda and Novak, Emily},
title = {Reliability and Agreement of Hand-Held Dynamometry Using Three Standard Rater Test Positions},
journal = {International Journal of Sports Physical Therapy},
year = {2025},
volume = {20},
number = {2},
doi = {10.26603/001c.128286}
}
EUVIMED