Purpose
Aerosol-generating procedures may increase infectious disease exposure, a particular concern during the early stages of the COVID-19 pandemic. This concern led to the development of various barrier devices. We developed a Negative Pressure Isolation Chamber (NPIC) to address the usability concerns of other devices. Our primary aim was to determine if the NPIC impedes intubation or diagnosis of an airway emergency whil limiting aerosol exposure.
Methods
With ethics approval, a within-subject, block-randomized evaluation study was conducted with anesthesiology staff, comparing timed intubation of a manikin with and without the NPIC. After successful intubation, a desaturation event was announced, requiring the removal of the NPIC if used. The primary outcome was intubation time. Secondary outcomes included time to diagnose an airway emergency, analysis of aerosols at the participant’s location, and subjective user experience with the NPIC.
Results
Twenty pediatric anesthesiologists participated. Use of the NPIC prolonged intubation time by a median 7.5 s [95% confidence interval (CI) 5.0 to 11.0 s; p < 0.001] and diagnosis of an emergency by 17.0 s [95%CI 15.0 to 19.0 s; p < 0.001]. However, aerosol concentration was lowered by 10.8 µg/m 3 (95%CI 7.7 to 14.0) when the NPIC was used. Eight (40%) participants deemed intubation more difficult with the NPIC. Feedback included concerns about obstructed video laryngoscope views and chamber ergonomics.
Conclusion
The NPIC delayed the time to intubation and diagnosis of an airway emergency by durations that may be of clinical significance. However, the NPIC also reduced participant aerosol particle exposure during preoxygenation, laryngoscopy and intubation.
Zitieren
APA
Baldelli, A., Poznikoff, A., West, N., Bone, J. N., Görges, M., & Purdy., R. (2026). Negative Pressure Isolation Chamber for Aerosol Generating Procedures: A Randomized Feasibility Simulation Trial. Anesthesia and Critical Care. https://doi.org/10.26502/acc.0112
Vancouver
Baldelli A, Poznikoff A, West N, Bone JN, Görges M, Purdy. R. Negative Pressure Isolation Chamber for Aerosol Generating Procedures: A Randomized Feasibility Simulation Trial. Anesthesia and Critical Care. 2026. doi:10.26502/acc.0112
BibTeX
@article{Baldelli2026,
author = {Baldelli, Alberto and Poznikoff, Andrew and West, Nicholas and Bone, Jeffrey N. and Görges, Matthias and Purdy., Robert},
title = {Negative Pressure Isolation Chamber for Aerosol Generating Procedures: A Randomized Feasibility Simulation Trial},
journal = {Anesthesia and Critical Care},
year = {2026},
doi = {10.26502/acc.0112}
}
EUVIMED