Intraluminal stone in a PD catheter : third world case
Bulletin de la Dialyse à Domicile · 2018
DOI: 10.25796/bdd.v1i1.34 ↗Kurz gefasst
Fachartikel in der Zeitschrift Bulletin de la Dialyse à Domicile, 2018. Verfasst von Béatrice Champtiaux Dechamp und 2 weiteren Autor:innen.
Abstract (Originalsprache)
Only 2 cases of intraluminal peritoneal catheter lithiasis have been reported so far. A woman (68 years of age) on CAPD was admitted to hospital because of intermittent outflow obstruction of her peritoneal catheter. A stone was found in the catheter lumen. Its biochemical composition was the same as that of the 2 other cases previously described: calcium-hydroxycardonate-phosphate with an admixture of protein. The stone could not migrate through the intraperitoneal tip hole of the catheter because its size (3 x 4 mm) was greater than the catheter lumen (2.7 mm). That suggests in situ development, directly in the catheter lumen despite inflow-outflow dialysate daily regular rinsings. In the 3 cases, dialysate contained lactate and 1.75 mmol/L calcium.
Zitieren
APA
Dechamp, B. C., Mandart, L., & Durand, P. (2018). Intraluminal stone in a PD catheter : third world case. Bulletin de la Dialyse à Domicile. https://doi.org/10.25796/bdd.v1i1.34
Vancouver
Dechamp BC, Mandart L, Durand P. Intraluminal stone in a PD catheter : third world case. Bulletin de la Dialyse à Domicile. 2018. doi:10.25796/bdd.v1i1.34
BibTeX
@article{Dechamp2018,
author = {Dechamp, Béatrice Champtiaux and Mandart, Lise and Durand, Pierre-Yves},
title = {Intraluminal stone in a PD catheter : third world case},
journal = {Bulletin de la Dialyse à Domicile},
year = {2018},
doi = {10.25796/bdd.v1i1.34}
}
EUVIMED