Diagnostic Accuracy of a Fluorescence Imaging Device in Diabetic Wounds: A Pilot Study Using a Tissue Culture System
Wounds: a compendium of clinical research and practice · 2023 · 35 (7) · e218-e223
DOI: 10.25270/wnds/23002 ↗Kurz gefasst
Fachartikel in der Zeitschrift Wounds: a compendium of clinical research and practice, 2023. Verfasst von Do-Yoon Koo und 4 weiteren Autor:innen. Die Arbeit wurde bisher 3-mal zitiert (Crossref).
Abstract (Originalsprache)
Introduction
DFUs are challenging chronic wounds that are vulnerable to infections. A fluorescence imaging device was developed to detect bacterial presence in wounds. This device utilizes the principle that when illuminated by violet light, some bacteria emit red fluorescence and others, such as Pseudomonas aeruginosa, emit cyan fluorescence. Several studies have reported the accuracy of this device. However, to the best of the authors’ knowledge, no studies have examined the correlation between bacterial presence and tissue biopsy culture results in diabetic wounds.
Objective
This study aimed to investigate the diagnostic accuracy of a fluorescence imaging device using a tissue culture system. Materials and Methods. Thirty-five patients (48 wounds) were included. Wounds were sampled using tissue culture methods and photographed using the fluorescence imaging device. Culture outcomes were categorized into non-Pseudomonas bacterial, Pseudomonas bacterial, both bacterial, and no-growth groups. Image outcomes were categorized into red, cyan, both colors, and negative groups.
Results
For detecting the presence of bacteria, the fluorescence imaging device showed a sensitivity, specificity, PPV, and NPV of 64.1%, 55.6%, 86.2%, and 26.3%, respectively, with an accuracy of 62.5%. For P aeruginosa, the device showed a sensitivity, specificity, PPV, and NPV of 66.7%, 87.2%, 54.6%, and 91.9%, respectively, with an accuracy of 83.3%. For non-Pseudomonas bacteria, the device showed a sensitivity, specificity, PPV, and NPV of 43.8%, 62.5%, 70.0%, and 35.7%, respectively, with an accuracy of 50.0%.
Conclusion
The fluorescence imaging device can help to detect the bacterial bioburden; however, its accuracy may be lower than that reported in previous studies of diabetic wounds.
Zitieren
APA
Koo, D., Namgoong, S., Han, S., Dhong, E., & Jeong, S. (2023). Diagnostic Accuracy of a Fluorescence Imaging Device in Diabetic Wounds: A Pilot Study Using a Tissue Culture System. Wounds: a compendium of clinical research and practice, 35(7), e218-e223. https://doi.org/10.25270/wnds/23002
Vancouver
Koo D, Namgoong S, Han S, Dhong E, Jeong S. Diagnostic Accuracy of a Fluorescence Imaging Device in Diabetic Wounds: A Pilot Study Using a Tissue Culture System. Wounds: a compendium of clinical research and practice. 2023;35(7):e218-e223. doi:10.25270/wnds/23002
BibTeX
@article{Koo2023,
author = {Koo, Do-Yoon and Namgoong, Sik and Han, Seung-Kyu and Dhong, Eun-Sang and Jeong, Seong-Ho},
title = {Diagnostic Accuracy of a Fluorescence Imaging Device in Diabetic Wounds: A Pilot Study Using a Tissue Culture System},
journal = {Wounds: a compendium of clinical research and practice},
year = {2023},
volume = {35},
number = {7},
pages = {e218--e223},
doi = {10.25270/wnds/23002}
}
EUVIMED