Contamination Risk During Fecal Management Device Removal: An In vitro, Simulated Clinical Use Study
Wound Management & Prevention · 2019
DOI: 10.25270/wmp.2019.3.3037 ↗Kurz gefasst
Fachartikel in der Zeitschrift Wound Management & Prevention, 2019. Verfasst von Daniel G Metcalf und 5 weiteren Autor:innen.
Abstract (Originalsprache)
Fecal management devices (FMDs) are used to drain and contain fecal matter in incontinent, often acutely or critically ill patients to protect their skin as well as the environment from contamination. However, there is potential for contamination and resultant infection at various stages of FMD use. PURPOSE: This in vitro study was conducted to compare device removal factors and subsequent splash of simulated fecal matter of 3 different designs of FMDs using a simulated rectum. METHODS: A Universal Test Machine was used to automatically measure removal forces (in newtons [N]) and tube extensions as the FMDs were pulled from the simulated rectum by the machine. Splash distance and quantity were measured using a splash-capture cylinder and image analysis software. Each device was tested 3 times. Two-sample t tests were conducted to examine statistical differences in removal forces, removal extensions, and splash areas. RESULTS: The forces required to remove the FMDs from the simulated rectum were significantly lower for the device with a collapsible, donut-shaped retention balloon compared with the devices with a green, foldable, trumpet-shaped retention cuff and a foldable, spherical-shaped retention balloon (12.0 ± 0.3 N vs. 32.6 ± 4.3 N and 34.8 ± 3.1 N, respectively; P
Zitieren
APA
Metcalf, D. G., Tsai, L., Williams, C., Pritchard, D., Parsons, D., & Bowler, P. G. (2019). Contamination Risk During Fecal Management Device Removal: An In vitro, Simulated Clinical Use Study. Wound Management & Prevention. https://doi.org/10.25270/wmp.2019.3.3037
Vancouver
Metcalf DG, Tsai L, Williams C, Pritchard D, Parsons D, Bowler PG. Contamination Risk During Fecal Management Device Removal: An In vitro, Simulated Clinical Use Study. Wound Management & Prevention. 2019. doi:10.25270/wmp.2019.3.3037
BibTeX
@article{Metcalf2019,
author = {Metcalf, Daniel G and Tsai, Lawrence and Williams, Corrine and Pritchard, David and Parsons, David and Bowler, Philip G},
title = {Contamination Risk During Fecal Management Device Removal: An In vitro, Simulated Clinical Use Study},
journal = {Wound Management & Prevention},
year = {2019},
doi = {10.25270/wmp.2019.3.3037}
}