Comparative analysis of antimicrobial susceptibility testing: Conventional versus direct from the blood culture bottle
Journal of Laboratory Physicians · 06.05.2026 · 18 · 342-347
DOI: 10.25259/jlp_337_2025 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Laboratory Physicians, online erschienen am 06.05.2026. Verfasst von Krishna Sarkar und 4 weiteren Autor:innen. Die Arbeit nennt 12 Literaturangaben.
Abstract (Originalsprache)
Objectives
The objective of the study is to evaluate the concordance, turnaround time (TAT), and clinical implications of direct antimicrobial susceptibility testing (AST) performed from positive BacT/ALERT ® blood culture bottles compared with conventional AST. Materials and Methods: A prospective study was conducted in the Microbiology Department of UCMS and GTB Hospital, including 250 positive blood cultures from patients with suspected bloodstream infections. Direct AST and conventional AST were performed according to Clinical & Laboratory Standards Institute (CLSI) guidelines. Categorical agreement (CA), TAT, and error rates were assessed. Statistical analysis: Descriptive statistics were used. Continuous variables (TAT) were compared using the Student’s t -test. CA between methods was evaluated using the Chi-square test with degrees of freedom and p values reported.
Results
Enterobacterales showed 94.4-99.4% CA, and non-fermenters demonstrated 92.8-100% agreement, with all error rates within CLSI-acceptable limits. Direct AST reduced mean TAT from 28 h (range: 18-32 h) to 9.7 h (range: 8-12.5 h). Direct AST demonstrated the greatest reduction in turnaround time in bloodstream and respiratory infection samples.
Conclusions
Direct AST demonstrated high concordance and significantly reduced TATs compared with conventional AST. It enables earlier targeted therapy and supports antimicrobial stewardship.
Zitieren
APA
Sarkar, K., Nirmal, K., Jayaraj, H., Banik, D., & Das, S. (2026). Comparative analysis of antimicrobial susceptibility testing: Conventional versus direct from the blood culture bottle. Journal of Laboratory Physicians, 18, 342-347. https://doi.org/10.25259/jlp_337_2025
Vancouver
Sarkar K, Nirmal K, Jayaraj H, Banik D, Das S. Comparative analysis of antimicrobial susceptibility testing: Conventional versus direct from the blood culture bottle. Journal of Laboratory Physicians. 2026;18:342-347. doi:10.25259/jlp_337_2025
BibTeX
@article{Sarkar2026,
author = {Sarkar, Krishna and Nirmal, Kirti and Jayaraj, Haarish and Banik, Debamita and Das, Shukla},
title = {Comparative analysis of antimicrobial susceptibility testing: Conventional versus direct from the blood culture bottle},
journal = {Journal of Laboratory Physicians},
year = {2026},
volume = {18},
pages = {342--347},
doi = {10.25259/jlp_337_2025}
}
EUVIMED