Proof of Concept Clinical Evaluation of Gradient Different Weight Threshold Segmentation for Accurate Tumour Delineation in I-131 SPECT Imaging
Indian Journal of Nuclear Medicine · 08.06.2026 · 41 · 284-297
DOI: 10.25259/ijnm_9_2026 ↗Kurz gefasst
Fachartikel in der Zeitschrift Indian Journal of Nuclear Medicine, online erschienen am 08.06.2026. Verfasst von Mohd Akmal Masud und 2 weiteren Autor:innen. Die Arbeit nennt 29 Literaturangaben.
Abstract (Originalsprache)
Objectives
We developed a modified gradient-different-weight threshold method, adapted from gradient-weighted thresholding, to improve robustness over fixed-percentage approaches. Material and Methods: The method computes voxel-wise gradient weights and optimises a threshold value by calibrating against a NEMA IQ phantom (six spheres) across sphere-to-background ratios (12:1, 10:1, 8:1, 6:1), linking the optimised threshold to a ratio factor (Cmax/Cmean) while comparing three background-estimation strategies (multi-circle, ULWL-guided narrow annular, and full wave with half maximum [FWHM]-based). Phantom optimisation showed the ULWL-guided narrow annular strategy produced the most consistent threshold-versus-ratio relationship. A proof-of-concept clinical evaluation on two retrospective I-131 SPECT/CT datasets used reference volumes from two nuclear medicine physicians (average contour) and reported percentage error, RMSD, and Dice similarity coefficient (DSC).
Results
For a thyroid-only case, the method yielded 27.1 mL versus 25.3 mL (9.71% error; DSC = 0.9541). In a whole-body multi-lesion case, delineation became less reliable when uptake was diffuse, lesions were low-count, or lesions were closely spaced, which elevated local background estimates.
Conclusion
These results suggest gradient-different-weight thresholding can improve I-131 SPECT tumour segmentation when background selection is feasible. However, larger multi-patient studies and direct side-by-side benchmarking are required to define its clinical operating range.
Zitieren
APA
Masud, M. A., Ngali, M. Z., & Said, M. A. (2026). Proof of Concept Clinical Evaluation of Gradient Different Weight Threshold Segmentation for Accurate Tumour Delineation in I-131 SPECT Imaging. Indian Journal of Nuclear Medicine, 41, 284-297. https://doi.org/10.25259/ijnm_9_2026
Vancouver
Masud MA, Ngali MZ, Said MA. Proof of Concept Clinical Evaluation of Gradient Different Weight Threshold Segmentation for Accurate Tumour Delineation in I-131 SPECT Imaging. Indian Journal of Nuclear Medicine. 2026;41:284-297. doi:10.25259/ijnm_9_2026
BibTeX
@article{Masud2026,
author = {Masud, Mohd Akmal and Ngali, Mohd Zamani and Said, Mohamad Aminudin},
title = {Proof of Concept Clinical Evaluation of Gradient Different Weight Threshold Segmentation for Accurate Tumour Delineation in I-131 SPECT Imaging},
journal = {Indian Journal of Nuclear Medicine},
year = {2026},
volume = {41},
pages = {284--297},
doi = {10.25259/ijnm_9_2026}
}
EUVIMED