The Novel Technique of using Superb Microvascular Imaging to Determine Carotid Intima-media Thickness
American Journal of Sonography · 2018
DOI: 10.25259/ajs-40-2018 ↗Kurz gefasst
Fachartikel in der Zeitschrift American Journal of Sonography, 2018. Verfasst von Fatima Musarrat Hasan und 1 weiteren Autor:innen. Die Arbeit nennt 10 Literaturangaben.
Abstract (Originalsprache)
Objective
The objective of this study was to investigate the interobserver reliability when measuring the carotid intima media thickness (IMT) using superb microvascular imaging (SMI) and B-mode ultrasonography.
Methods
Two sonologists were selected to scan the left common carotid artery and measure IMT first with B-mode and then with SMI on 20 patients. They were blinded to each other results. Intraclass correlation coefficients (ICCs) were calculated to estimate the inter-rater reliability using both the modes of scanning.
Results
Interobserver agreement when using SMI, for both near wall and far wall, was almost perfect (ICC, 0.870; 95% confidence interval [CI], 0.700–0.946). Interobserver agreement when using B-mode was poor for near wall (ICC, 0.396; 95% CI, −0.048–0.708) and moderate for far wall (ICC, 0.474, 95% CI, 0.070–0.749).
Conclusions
SMI proved to be a greatly reliable tool in the measurement of carotid IMT.
Zitieren
APA
Hasan, F. M., & Hasan, M. (2018). The Novel Technique of using Superb Microvascular Imaging to Determine Carotid Intima-media Thickness. American Journal of Sonography. https://doi.org/10.25259/ajs-40-2018
Vancouver
Hasan FM, Hasan M. The Novel Technique of using Superb Microvascular Imaging to Determine Carotid Intima-media Thickness. American Journal of Sonography. 2018. doi:10.25259/ajs-40-2018
BibTeX
@article{Hasan2018,
author = {Hasan, Fatima Musarrat and Hasan, Musarrat},
title = {The Novel Technique of using Superb Microvascular Imaging to Determine Carotid Intima-media Thickness},
journal = {American Journal of Sonography},
year = {2018},
doi = {10.25259/ajs-40-2018}
}
EUVIMED