Advances in Central Visual Field Testing: A Narrative Review of the 24-2C SITA Faster Strategy

Shruti Jha, Pinaz Nasim

International Journal of Drug Delivery Technology · 11.06.2026 · 16 (52s)

DOI: 10.25258/ijddt.16.52s.40 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Drug Delivery Technology, online erschienen am 11.06.2026. Verfasst von Shruti Jha und 1 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Standard automated perimetry remains a cornerstone in the diagnosis and monitoring of glaucoma. The conventional 24-2 test grid has been widely adopted in clinical practice due to its balance between spatial coverage and test duration, but it provides limited sampling of the central visual field. As the need to examine changes in the central visual field increases, the 24-2C grid has been developed to bridge this gap by incorporating additional central test locations into the existing 24-2 framework. This modification aims to improve the detection of central defects without substantially increasing testing time.

Purpose

To synthesize current evidence on the clinical utility of the 24-2C SITA Faster strategy for central visual field (VF) assessment and to compare its performance with conventional 24-2 and 10-2 perimetric grids.

Methods

A narrative review of peer-reviewed studies was conducted using PubMed, Scopus, and Google Scholar up to 2025. Studies reporting outcomes on central defect detection, global indices, diagnostic accuracy, structure–function relationships, reliability, and test duration were included.

Results

The 24-2C grid improves detection of central defects compared to 24-2, identifying additional defective points and clusters within the central 10°. Differences in global indices (MD, PSD) are small and often not significant with SITA Faster. Compared with 10-2, 24-2C demonstrates lower sensitivity for fine paracentral defects but shows moderateto-strong agreement in global indices. Diagnostic accuracy is improved with 24-2C central sampling, and test duration remains shorter than 24-2 and10-2 grids.

Conclusion

The 24-2C SITA Faster strategy enhances central VF assessment and offers a practical balance between diagnostic yield and efficiency. It should be used as part of a tailored testing approach rather than replacing 10-2. Clinical Significance: This narrative review provides a consolidated appraisal of current evidence on the 24-2C SITA Faster strategy, offering clinicians a structured understanding of its role in central visual field assessment. By integrating findings across comparative, diagnostic, and structure–function studies, it clarifies the relative strengths and limitations of 24-2C in relation to conventional 24-2 and 10-2 strategies.

Zitieren

APA

Jha, S., & Nasim, P. (2026). Advances in Central Visual Field Testing: A Narrative Review of the 24-2C SITA Faster Strategy. International Journal of Drug Delivery Technology, 16(52s). https://doi.org/10.25258/ijddt.16.52s.40

Vancouver

Jha S, Nasim P. Advances in Central Visual Field Testing: A Narrative Review of the 24-2C SITA Faster Strategy. International Journal of Drug Delivery Technology. 2026;16(52s). doi:10.25258/ijddt.16.52s.40

BibTeX
@article{Jha2026,
  author = {Jha, Shruti and Nasim, Pinaz},
  title = {Advances in Central Visual Field Testing: A Narrative Review of the 24-2C SITA Faster Strategy},
  journal = {International Journal of Drug Delivery Technology},
  year = {2026},
  volume = {16},
  number = {52s},
  doi = {10.25258/ijddt.16.52s.40}
}
Kostenlos starten →