Comparative Study of LMA Blockbuster and i-gel Airway Devices for Hemodynamic Stability in Patients Undergoing General Anesthesia
International Journal of Drug Delivery Technology · 01.06.2026 · 16 (39s)
DOI: 10.25258/ijddt.16.39s.44 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Drug Delivery Technology, erschienen am 01.06.2026. Verfasst von Rhutu Dakhole und 5 weiteren Autor:innen.
Abstract (Originalsprache)
Background
Supraglottic airway devices are widely used during general anaesthesia. The i-gel with a non-inflatable gel cuff and the newer BLOCKBUSTER with a low pressure silicone cuff are both second generation devices, but direct comparison of their haemodynamic profiles and patient centred outcomes is limited.
Aim
To compare the haemodynamic stability, insertion characteristics, postoperative complications, and patient comfort of i-gel and Blockbuster™ in adult patients undergoing elective surgery under general anaesthesia.
Methods
In this cross-sectional comparative study, 70 adults (ASA I–II) were allocated equally to i-gel or BLOCKBUSTER. The primary outcome was heart rate, blood pressure, and SpO₂ at baseline, immediately after insertion, at 1 minute, 3 minute, and 5 minute. Secondary outcomes included insertion time, first attempt success, ease of insertion, sore throat, hoarseness, dysphagia, cough, and patient comfort score (1–10). Data were analysed using t test, Mann Whitney U, and chi square test (p0.05). No significant differences were observed in any haemodynamic parameter at any time point (p>0.05 for all). Insertion time, first attempt success, and ease of insertion were comparable (p=0.634, 0.643, 0.691, respectively). Postoperative sore throat (25.7% vs 34.3%), hoarseness, dysphagia, cough, and comfort scores (median 10 in both groups) did not differ significantly (p>0.05 for all).
Conclusion
Both i-gel and BLOCKBUSTER provide equivalent haemodynamic stability, ease of use, and patient comfort. Device selection may be guided by availability and cost rather than clinical superiority.
Zitieren
APA
Dakhole, R., Singh, A. V., ., R., Singh, K., Londhe, P., & Viswambharan, A. (2026). Comparative Study of LMA Blockbuster and i-gel Airway Devices for Hemodynamic Stability in Patients Undergoing General Anesthesia. International Journal of Drug Delivery Technology, 16(39s). https://doi.org/10.25258/ijddt.16.39s.44
Vancouver
Dakhole R, Singh AV, . R, Singh K, Londhe P, Viswambharan A. Comparative Study of LMA Blockbuster and i-gel Airway Devices for Hemodynamic Stability in Patients Undergoing General Anesthesia. International Journal of Drug Delivery Technology. 2026;16(39s). doi:10.25258/ijddt.16.39s.44
BibTeX
@article{Dakhole2026,
author = {Dakhole, Rhutu and Singh, Ajay Vikram and ., Rachna and Singh, Kulvir and Londhe, Prerana and Viswambharan, Akshaya},
title = {Comparative Study of LMA Blockbuster and i-gel Airway Devices for Hemodynamic Stability in Patients Undergoing General Anesthesia},
journal = {International Journal of Drug Delivery Technology},
year = {2026},
volume = {16},
number = {39s},
doi = {10.25258/ijddt.16.39s.44}
}
EUVIMED